Abstract
Imaging fibre bundles and a conical mirror have been used to construct a passive, endoscopic OCT probe with no scanning components at the probe tip. Circular scanning of the beam projected onto the proximal face of the imaging bundle results in a corresponding circular scan at the distal end of the probe. A 45° base-angle conical mirror turns the output light to produce a radially-directed beam that permits circumferential OCT scanning in quasi-cylindrical ducts.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Helen D. Ford, Helen D. Ford, Ralph P. Tatam, Ralph P. Tatam, "Optical coherence tomography for endoscopes, using imaging fibre bundles and a conical mirror", Proc. SPIE 8421, OFS2012 22nd International Conference on Optical Fiber Sensors, 84215N (7 November 2012); doi: 10.1117/12.974499; https://doi.org/10.1117/12.974499