13 September 2012 Optical calibration of capacitive sensors for AO: strategy and preliminary results
Author Affiliations +
Abstract
LBT contactless, voice-coil actuators working point ranges from 10 to 100 μm, with a typical position noise of a few nm. To fully take advantage of such large working range in AO operations, an absolute calibration of their capacitive sensor is requested. We implemented and tested an optical calibration technique that is based on a fast interferometer detection of a differential piston superimposed to a low order mirror mode. This procedure has been tested in dome environment at the Large Binocular Telescope.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Runa Briguglio, Marco Xompero, Armando Riccardi, Roberto Biasi, Mario Andrighettoni, "Optical calibration of capacitive sensors for AO: strategy and preliminary results", Proc. SPIE 8447, Adaptive Optics Systems III, 84474E (13 September 2012); doi: 10.1117/12.926820; https://doi.org/10.1117/12.926820