13 September 2012 Unintended consequences on vacuum hold time with conversion to PolyCold closed-cycle coolers
Author Affiliations +
Abstract
In 2009, the Canada-France-Hawaii Telescope converted the dewar for the ESPaDOnS spectrograph from using liquid nitrogen cooling to using a Polycold® closed-cycle cooler. It was found that the higher temperature of the closed-cycle cooler degraded the efficiency of the carbon getter for maintaining the vacuum to the point that vacuum could not be maintained long-term without the pumping of a cold-cathode gauge. This paper will detail this vacuum issue along with some experiments done to show the pumping rates of the carbon getter and cold-cathode vacuum gauges, and describe a light-baffle design for suppressing light from the gauge.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tom Benedict, Tom Benedict, Gregory A. Barrick, Gregory A. Barrick, "Unintended consequences on vacuum hold time with conversion to PolyCold closed-cycle coolers", Proc. SPIE 8450, Modern Technologies in Space- and Ground-based Telescopes and Instrumentation II, 84504M (13 September 2012); doi: 10.1117/12.926428; https://doi.org/10.1117/12.926428