Paper
17 October 2012 NPS assessment of color medical image displays using a monochromatic CCD camera
Author Affiliations +
Abstract
This paper presents an approach to Noise Power Spectrum (NPS) assessment of color medical displays without using an expensive imaging colorimeter. The R, G and B color uniform patterns were shown on the display under study and the images were taken using a high resolution monochromatic camera. A colorimeter was used to calibrate the camera images. Synthetic intensity images were formed by the weighted sum of the R, G, B and the dark screen images. Finally the NPS analysis was conducted on the synthetic images. The proposed method replaces an expensive imaging colorimeter for NPS evaluation, which also suggests a potential solution for routine color medical display QA/QC in the clinical area, especially when imaging of display devices is desired
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hans Roehrig, Xiliang Gu, and Jiahua Fan "NPS assessment of color medical image displays using a monochromatic CCD camera", Proc. SPIE 8508, Medical Applications of Radiation Detectors II, 85080C (17 October 2012); https://doi.org/10.1117/12.966331
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cameras

LCDs

CCD cameras

Imaging devices

Calibration

Medical imaging

Image quality

Back to Top