20 November 2012 Simultaneous measurements of atmospheric NO2 and HONO using IBBCEAS with a near-ultraviolet LED
Author Affiliations +
Abstract
High sensitivity simultaneous measurements of atmospheric NO2 and HONO using incoherent broadband cavityenhanced absorption spectroscopy (IBBCEAS) were developed. A near-ultraviolet light emitting diode (LED) peaked at 372 nm was used as light source of the IBBCEAS instrument to measure the absorption of NO2 and HONO in the spectral range of 361-378 nm. Concentrations of atmospheric NO2 and HONO were retrieved from the absorption spectra recorded in the absence of atmospheric aerosols and the average detection limits of 2.9 ppbv for NO2 and 1.2 ppbv for HONO with an acquisition time of 5 min were achieved. The results demonstrated high sensitivity of this measurement technique based on IBBCEAS, which is a promising technique for measurements of atmospheric trace gases.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Liuyi Ling, Liuyi Ling, Pinhua Xie, Pinhua Xie, Min Qin, Min Qin, Renzhi Hu, Renzhi Hu, Nina Zheng, Nina Zheng, Fuqi Si, Fuqi Si, } "Simultaneous measurements of atmospheric NO2 and HONO using IBBCEAS with a near-ultraviolet LED", Proc. SPIE 8563, Optical Metrology and Inspection for Industrial Applications II, 856310 (20 November 2012); doi: 10.1117/12.999572; https://doi.org/10.1117/12.999572
PROCEEDINGS
7 PAGES


SHARE