Paper
20 March 2013 Effects of sterilization methods on key properties of specialty optical fibers used in medical devices
Andrei A. Stolov, Brian E. Slyman, David T. Burgess, Adam S. Hokansson, Jie Li, R. Steve Allen
Author Affiliations +
Abstract
Optical fibers with different types of polymer coatings were exposed to three sterilization conditions: multiple autoclaving, treatment with ethylene oxide and treatment with gamma rays. Effects of different sterilization techniques on key optical and mechanical properties of the fibers are reported. The primary attention is given to behavior of the coatings in harsh sterilization environments. The following four coating/buffer types were investigated: (i) dual acrylate, (ii) polyimide, (iii) silicone/PEEK and (iv) fluoroacrylate hard cladding/ETFE.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Andrei A. Stolov, Brian E. Slyman, David T. Burgess, Adam S. Hokansson, Jie Li, and R. Steve Allen "Effects of sterilization methods on key properties of specialty optical fibers used in medical devices", Proc. SPIE 8576, Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications XIII, 857606 (20 March 2013); https://doi.org/10.1117/12.1000005
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical fibers

Gamma radiation

Silicon

Polymers

Signal attenuation

Cladding

Neodymium

RELATED CONTENT

Ormosil coatings for POF
Proceedings of SPIE (February 09 1993)
UV curable low index hybrid glass as hard cladding for...
Proceedings of SPIE (September 23 2005)
Optical fiber coatings for medical applications
Proceedings of SPIE (August 01 1992)

Back to Top