Abstract
We study multi-shot intensity-and-phase measurements of unstable trains of ultrashort pulses using spectral-phase interferometry for direct electric-field reconstruction (SPIDER), second harmonic generation (SHG) frequency-resolved optical gating (FROG), polarization gate (PG) FROG, and cross-correlation FROG (XFROG). An analytical calculation suggests that SPIDER cannot indicate instability in pulse trains well. Simulations confirm this and demonstrate that SPIDER only measures the coherent artifact. Further, the presence of instability cannot be distinguised from benign misalignment effects in SPIDER. FROG methods suggest instability by exhibiting clear disagreement between measured and retrieved traces.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Michelle Rhodes, Günter Steinmeyer, Justin Ratner, and Rick Trebino "The coherent artifact in modern pulse measurement", Proc. SPIE 8611, Frontiers in Ultrafast Optics: Biomedical, Scientific, and Industrial Applications XIII, 86110R (15 March 2013); https://doi.org/10.1117/12.2003302
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Second-harmonic generation

Neodymium

Visibility

Ions

Mode locking

Reliability

Interferometry