Paper
6 March 2013 Demonstration of a scatter correction technique in digital breast tomosynthesis
Christy R. Inscoe, Andrew W. Tucker, Otto Z. Zhou, Jianping Lu
Author Affiliations +
Proceedings Volume 8668, Medical Imaging 2013: Physics of Medical Imaging; 86680H (2013) https://doi.org/10.1117/12.2008044
Event: SPIE Medical Imaging, 2013, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
We have recently developed a method of using a distributed x-ray source array to obtain images with scatter correction for tomographic reconstruction of an object. The method consists of obtaining x-ray images of the object with and without the primary beam sampling apparatus. In this study, we report the results of applying the scatter correction method for breast tomosynthesis imaging using the carbon nanotube x-ray based stationary Digital Breast Tomosynthesis (s-DBT) system developed at UNC. The unique design of s-DBT system makes it possible to estimate the image of the scatter profile of the object with very low dose, and without significant increase in acquisition time. An anthropomorphic breast phantom was used for quantitative analysis of the change in contrast and scatter-to-primary ratio. Our results suggest that the scatter correction method is effective and can be used for enhanced contrast.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Christy R. Inscoe, Andrew W. Tucker, Otto Z. Zhou, and Jianping Lu "Demonstration of a scatter correction technique in digital breast tomosynthesis", Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86680H (6 March 2013); https://doi.org/10.1117/12.2008044
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Breast

Digital breast tomosynthesis

Image processing

X-ray sources

X-rays

Computer aided design

3D image reconstruction

Back to Top