6 March 2013 Optimizing the acquisition parameters of a newly developed digital breast tomosynthesis system
Hye-Suk Park, Ye-Seul Kim, JaeGu Choi, Young-Wook Choi, Hee-Joung Kim
Author Affiliations +
Proceedings Volume 8668, Medical Imaging 2013: Physics of Medical Imaging; 86684F (2013) https://doi.org/10.1117/12.2007453
Event: SPIE Medical Imaging, 2013, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
The purpose of this study was to investigate the effect of different acquisition parameters and to characterize their relationships in order to improve the detection of microcalcifications using digital breast tomosynthesis (DBT). DBT imaging parameters were optimized using 32 different acquisition sets with six angular ranges (±5°, ±10°, ±13°, ±17°, ±21°, and ±25°) and eight projection views (5, 11, 15, 21, 25, 31, 41, and 51 projections). To investigate the effects of variable angular dose distribution, the acquisition sets were evaluated with delivering more dose toward the central views. Our results show that a wide angular range improved the reconstructed image quality in the z-direction. If a large number of projections are acquired, then electronic noise may dominate the contrast-to-noise ratio (CNR) due to reduced radiation dose per projection. With delivering more dose toward the central views, it was found that the vertical resolution was reduced with increasing dose in the central PVs. On the other hand, the CNR clearly increased with increasing concentration of dose distribution in central views. Although it was found that increasing angular range improved the vertical resolution, it was also found that the image quality of microcalcifications in the in-focus plane did not improve by increasing the noise due to greater effective breast thickness. Angular dose distributions, with more dose delivered to the central views, generally yielded a higher quality factor (QF) than uniform dose distributions.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hye-Suk Park, Ye-Seul Kim, JaeGu Choi, Young-Wook Choi, and Hee-Joung Kim "Optimizing the acquisition parameters of a newly developed digital breast tomosynthesis system", Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86684F (6 March 2013); https://doi.org/10.1117/12.2007453
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Digital breast tomosynthesis

Image quality

Breast

Sensors

Image acquisition

Photovoltaics

Reconstruction algorithms

Back to Top