You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
19 March 2013Preliminary study on phase-contrast digital tomosynthesis:
development and evaluation of experimental system
The advantage of X-ray phase imaging is its ability to obtain information on soft tissues, which is difficult
using conventional X-ray imaging. Moreover, a sharp X-ray image can be obtained from the edge effect
resulting from phase contrast. Digital tomosynthesis is an imaging technique used to reconstruct multiple
planes in a single scan. In this study, we developed an experimental system that combines the
phase-contrast and digital tomosynthesis techniques. Our experimental system consists of a
transmission-type micro-focus X-ray source (minimum focus size: 1 μm). We also introduced an indirect
conversion-type flat panel detector (pixel pitch: 50 μm, matrix size: 2366 × 2368) as an imaging device.
The sample is placed on a computer-controlled rotation table, and projection images are captured from
various angles. The images are then reconstructed using the filtered back projection method. In the
experiments, a tomosynthesis image of an acrylic phantom was obtained at a tube voltage of 40 kV and at
a maximum projection angle of ±20°. To evaluate the edge enhancement effect by phase contrast, the
resolution, degree of edge enhancement, and image contrast were measured using the acrylic phantom. A
good edge enhancement effect was confirmed under the specified conditions. Furthermore, we compared
to the shape between the projection image and the tomosynthesis image and found that the tomosynthesis
image showed high shape reproducibility compared to the conventional projection image. These results
indicate that phase-contrast digital tomosynthesis may be useful for the three-dimensional imaging of
low-contrast material.
The alert did not successfully save. Please try again later.
Ai Ikeya, Atsushi Teramoto, Kenji Noguchi, Hiroshi Fujita, "Preliminary study on phase-contrast digital tomosynthesis:
development and evaluation of experimental system," Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 866853 (19 March 2013); https://doi.org/10.1117/12.2007855