14 March 2013 Freehand ultrasound calibration: phantom versus tracked pointer
Author Affiliations +
Abstract
PURPOSE: Ultrasound-guided tracked navigation requires spatial calibration between the ultrasound beam and the tracker. We examined the reproducibility and accuracy of two popular open source calibration methods1 with a handheld linear ultrasound transducer. METHODS: A total of 10 calibrations were performed using (1) a double N-wire phantom with automatic image segmentation and registration; (2) and registration of landmark points collected with a tracked pointer. Reproducibility and accuracy were characterized by comparing the resulting transformation matrices, and by comparing ground truth landmark points. RESULTS: Transformation matrices calculated with an N-wire phantom showed a variance of X: 0.02 mm (in the direction of sound propagation), Y: 0.03 mm (in the direction of transducer elements) and Z: 0.21 mm (in the elevation direction). Transformation matrices obtained with tracked pointer showed a variance of X: 0.1 mm, Y: 0.10 mm and Z: 0.43 mm. Calibration accuracy was tested with ground truth cross wire points. The N-wire phantom provided a calibration with a distance from ground truth of X: 2.44 ± 1.44 mm, Y: 1.21 ± 0.88 mm, and Z: 1.12 ± 0.82 mm. Tracked pointer calibration had a distance from the ground truth of X: 0.23 ± 0.16 mm, Y: 0.62 ± 0.31 mm, and Z: 0.45 ± 0.33 mm. Distance from ground truth was significantly less (p<0.01) with the tracked pointer method in all directions. CONCLUSION: Calibration using a tracked pointer had a slightly greater variance; however it showed better accuracy over calibrations calculated with N-wire phantoms.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mattea Welch, Mattea Welch, Jennifer Andrea, Jennifer Andrea, Tamas Ungi, Tamas Ungi, Gabor Fichtinger, Gabor Fichtinger, "Freehand ultrasound calibration: phantom versus tracked pointer", Proc. SPIE 8671, Medical Imaging 2013: Image-Guided Procedures, Robotic Interventions, and Modeling, 86711C (14 March 2013); doi: 10.1117/12.2007922; https://doi.org/10.1117/12.2007922
PROCEEDINGS
6 PAGES


SHARE
Back to Top