29 March 2013 3D ultrasound imaging performance of a row-column addressed 2D array transducer: a simulation study
Morten Fischer Rasmussen, Jørgen Arendt Jensen
Author Affiliations +
Abstract
This paper compares the imaging performance of a 128+128 element row-column addressed array with a fully addressed 16×16 2D array. The comparison is made via simulations of the point spread function with Field II. Both arrays have lambda-pitch, a center frequency of 3:5MHz and use 256 active elements. The row-column addressed array uses 128 transmit channels and 128 receive channels, whereas the fully addressed array uses 256 channels in both transmit and receive. The large size of the emulated row and column elements in the row-column addressed array causes ghost echoes to appear. The ghost echoes are shown to be suppressed when the sub-elements within each of the emulated row and column elements are apodized. The maximum ghost intensity is suppressed by 22:2 dB compared to using no apodization. With apodization applied, the full-width-at-half-maximum in the lateral direction for the fully addressed array is 2:81mm, and 1:01mm for the row-column addressed array. This shows that the detail resolution can be more than doubled using the row-column addressed array instead of the fully addressed array. The row column addressed array achieves a R20 dB cystic resolution of 0:76mm, compared to 3:16mm for the fully addressed array. The significantly smaller R20 dB-value for the row-column addressed array indicates that it can achieve a much higher contrast resolution than the fully addressed array.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Morten Fischer Rasmussen and Jørgen Arendt Jensen "3D ultrasound imaging performance of a row-column addressed 2D array transducer: a simulation study", Proc. SPIE 8675, Medical Imaging 2013: Ultrasonic Imaging, Tomography, and Therapy, 86750C (29 March 2013); https://doi.org/10.1117/12.2007083
Lens.org Logo
CITATIONS
Cited by 19 scholarly publications and 11 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Apodization

Chemical elements

Point spread functions

Transducers

Ultrasonography

Imaging arrays

3D image processing

Back to Top