Translator Disclaimer
22 June 2013 A numerical method for designing gradient-index fiber probes
Author Affiliations +
Proceedings Volume 8769, International Conference on Optics in Precision Engineering and Nanotechnology (icOPEN2013); 87693B (2013) https://doi.org/10.1117/12.2017818
Event: International Conference on Optics in Precision Engineering and Nanotechnology (icOPEN2013), 2013, Singapore, Singapore
Abstract
A numerical method is investigated to design gradient-index (GRIN) fiber probes. The GRIN fiber probe is composed of a single mode fiber (SMF), a no-core fiber (NCF), and a GRIN fiber lens. The optical software GLAD is adopted to simulate the optical performance of the probe. The simulation results show that, given the length of the GRIN fiber lens 0.1mm and the length of the NCF 0.36mm, the working distance is 0.73mm and the focus spot size 33μm, which are well agreement with the experimental data. As a result, the proposed numerical method is validated to be effective to design such GRIN fiber probes.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Chi Wang, Xueqin Xia, Shubo Bi, Xingfei Li, Li Wang, and Yingjie Yu "A numerical method for designing gradient-index fiber probes", Proc. SPIE 8769, International Conference on Optics in Precision Engineering and Nanotechnology (icOPEN2013), 87693B (22 June 2013); https://doi.org/10.1117/12.2017818