18 November 2013 System for measuring the angular response of radiometers
Author Affiliations +
Proceedings Volume 8785, 8th Iberoamerican Optics Meeting and 11th Latin American Meeting on Optics, Lasers, and Applications; 87852F (2013) https://doi.org/10.1117/12.2019888
Event: 8th Ibero American Optics Meeting/11th Latin American Meeting on Optics, Lasers, and Applications, 2013, Porto, Portugal
Abstract
To measure the angular response of radiometers developed in the laboratory, an automated system was assembled and characterized. The main light source is a QTH lamp of 1000W and for UV radiometers a Hg(Xe) lamp of 500W. The light beam has a useful diameter of 40 mm, a half divergence of 10mrad and the spatial uniformity is better than 97%. Errors up to 5% at large angles were estimated for misalignments of 1° in the positioning of the radiometers . In this work details of the system and results of measurements obtained with radiometers developed in the laboratory and with commercial ones are presented.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Luiz Ângelo Berni, Waldeir Amaral Vilela, Antonio Fernando Beloto, Felipe Oliveira de Sena, "System for measuring the angular response of radiometers", Proc. SPIE 8785, 8th Iberoamerican Optics Meeting and 11th Latin American Meeting on Optics, Lasers, and Applications, 87852F (18 November 2013); doi: 10.1117/12.2019888; https://doi.org/10.1117/12.2019888