14 June 2013 Optical attenuation correction in multispectral optoacoustic tomography with logarithm unmixing
Author Affiliations +
Abstract
Quantification of extrinsically administered contrast agents in optoacoustic (photoacoustic) tomography is a challenging task, mainly due to spectrally-dependent contributions from absorbing background tissue chromophores leading to strong changes in the light fluence for different positions and wavelengths. Herein we present a procedure capable of self-calibrating light fluence variations for quantitative imaging of the distribution of photo-absorbing agents. The method makes use of a logarithmic representation of the images taken at different wavelengths assisted with a blind unmixing approach. It is shown that the serial expansion of the logarithm of an image contains a term representing the ratio between absorption of the probe of interest and other background components. Provided the background variations are not very high, this term can be isolated with an unmixing algorithm, so that the concentration of the probe can subsequently be resolved.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
X. Luís Deán-Ben, X. Luís Deán-Ben, Andreas Buehler, Andreas Buehler, Vasilis Ntziachristos, Vasilis Ntziachristos, Daniel Razansky, Daniel Razansky, } "Optical attenuation correction in multispectral optoacoustic tomography with logarithm unmixing", Proc. SPIE 8800, Opto-Acoustic Methods and Applications, 88000H (14 June 2013); doi: 10.1117/12.2033351; https://doi.org/10.1117/12.2033351
PROCEEDINGS
5 PAGES


SHARE
Back to Top