18 February 2014 Er:YAG laser delivery sytems and sonic-activated bulk composite restoration: sculpturing and microleakage evaluation
Author Affiliations +
Proceedings Volume 8929, Lasers in Dentistry XX; 892909 (2014) https://doi.org/10.1117/12.2036771
Event: SPIE BiOS, 2014, San Francisco, California, United States
Abstract
The comparison of tissue quality and its restoration after contact, non-contact, and non-contact scanned Er:YAG (2940 nm) laser radiation ablation was evaluated. Laser setting for contact ablation was 250 mJ/pulse, pulse repetition rate 15 Hz, average power 3.75 W. For non-contact ablation these values were 600 mJ/pulse, 6 Hz, 3.6 W. The scanning ablation was provided in non-contact mode (1440 pulses/1 cavity). All cavities were filled by sonic activated composite resin. Microleakage was assessed quantitatively by the degree of dye penetration. Cavity prepared by contact mode with sonic-activated composite filling protects microleakage formation.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tatjána Dostálová, Helena Jelínková, Jan Šulc, Michal Němec, Michaela Bučková, Magdalena Kašparová, Pavel Bradna, "Er:YAG laser delivery sytems and sonic-activated bulk composite restoration: sculpturing and microleakage evaluation", Proc. SPIE 8929, Lasers in Dentistry XX, 892909 (18 February 2014); doi: 10.1117/12.2036771; https://doi.org/10.1117/12.2036771