3 March 2014 Development of a widefield phantom eye for retinal optical coherence tomography
Author Affiliations +
Abstract
We report the design, manufacture and assessment of a phantom eye that can be used to measure the performance and accuracy of ophthalmic-OCT devices. We base our design on a wide-field schematic eye, R. Navarro, J. Opt. Soc. Am. A 2 (1985), to allow the assessment of device performance relative to ± 70° external field of view. We have fabricated the phantom eye and have verified the structural dimensions of the multi-material 3D-printed retinal targets using calibrated-OCT images.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Anthony T. Corcoran, Gonzalo Muyo, Jano I. van Hemert, Andrew R. Harvey, "Development of a widefield phantom eye for retinal optical coherence tomography", Proc. SPIE 8945, Design and Performance Validation of Phantoms Used in Conjunction with Optical Measurement of Tissue VI, 89450F (3 March 2014); doi: 10.1117/12.2037767; https://doi.org/10.1117/12.2037767