12 March 2014 Spectral-domain interferometry for quantitative DIC microscopy
Author Affiliations +
Abstract
A spectral-domain differential interference contrast (SD-DIC) microscopy system is presented for quantitative imaging of both reflective and transparent samples. The spectral-domain interferometry, combined with the common-path DIC geometry, provides a shot noise-limited sensitivity of 14.3pm in optical pathlength gradient measurement. The optical resolution of the system was characterized using images of a USAF resolution target. Fused silica microspheres were imaged to demonstrate the reconstruction of two-dimensional optical pathlength topography from measured gradient fields. The exquisite sensitivity of the system showed potential in quantitative imaging of sub-diffraction limit objects such as gold nanoparticles.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Chengshuai Li, Chengshuai Li, Yizheng Zhu, Yizheng Zhu, } "Spectral-domain interferometry for quantitative DIC microscopy", Proc. SPIE 8949, Three-Dimensional and Multidimensional Microscopy: Image Acquisition and Processing XXI, 89491D (12 March 2014); doi: 10.1117/12.2040541; https://doi.org/10.1117/12.2040541
PROCEEDINGS
5 PAGES


SHARE
Back to Top