You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
19 March 2014Experimental and theoretical performance analysis for a CMOS-based high resolution image detector
Increasing complexity of endovascular interventional procedures requires superior x-ray imaging quality. Present stateof-
the-art x-ray imaging detectors may not be adequate due to their inherent noise and resolution limitations. With recent
developments, CMOS based detectors are presenting an option to fulfill the need for better image quality. For this work,
a new CMOS detector has been analyzed experimentally and theoretically in terms of sensitivity, MTF and DQE.
The detector (Dexela Model 1207, Perkin-Elmer Co., London, UK) features 14-bit image acquisition, a CsI phosphor, 75
μm pixels and an active area of 12 cm x 7 cm with over 30 fps frame rate. This detector has two modes of operations
with two different full-well capacities: high and low sensitivity. The sensitivity and instrumentation noise equivalent
exposure (INEE) were calculated for both modes. The detector modulation-transfer function (MTF), noise-power spectra
(NPS) and detective quantum efficiency (DQE) were measured using an RQA5 spectrum. For the theoretical
performance evaluation, a linear cascade model with an added aliasing stage was used.
The detector showed excellent linearity in both modes. The sensitivity and the INEE of the detector were found to be
31.55 DN/μR and 0.55 μR in high sensitivity mode, while they were 9.87 DN/μR and 2.77 μR in low sensitivity mode.
The theoretical and experimental values for the MTF and DQE showed close agreement with good DQE even at
fluoroscopic exposure levels.
In summary, the Dexela detector’s imaging performance in terms of sensitivity, linear system metrics, and INEE
demonstrates that it can overcome the noise and resolution limitations of present state-of-the-art x-ray detectors.
The alert did not successfully save. Please try again later.
Amit Jain, Daniel R. Bednarek, Stephen Rudin, "Experimental and theoretical performance analysis for a CMOS-based high resolution image detector," Proc. SPIE 9033, Medical Imaging 2014: Physics of Medical Imaging, 90333P (19 March 2014); https://doi.org/10.1117/12.2043053