You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
19 March 2014A new resonance-frequency based electrical impedance spectroscopy and its application in biomedical engineering
Electrical Impedance Spectroscopy (EIS) has shown promising results for differentiating between malignant and
benign tumors, which exhibit different dielectric properties. However, the performance of current EIS systems has been
inadequate and unacceptable in clinical practice. In the last several years, we have been developing and testing a new
EIS approach using resonance frequencies for detection and classification of suspicious tumors. From this experience,
we identified several limitations of current technologies and designed a new EIS system with a number of new
characteristics that include (1) an increased A/D (analog-to-digital) sampling frequency, 24 bits, and a frequency
resolution of 100 Hz, to increase detection sensitivity (2) automated calibration to monitor and correct variations in
electronic components within the system, (3) temperature sensing and compensation algorithms to minimize impact of
environmental change during testing, and (4) multiple inductor-switching to select optimum resonance frequencies. We
performed a theoretical simulation to analyze the impact of adding these new functions for improving performance of the
system. This system was also tested using phantoms filled with variety of liquids. The theoretical and experimental test
results are consistent with each other. The experimental results demonstrated that this new EIS device possesses the
improved sensitivity and/or signal detection resolution for detecting small impedance or capacitance variations. This
provides the potential of applying this new EIS technology to different cancer detection and diagnosis tasks in the future.