12 March 2014 Solving for free-hand and real-time 3D ultrasound calibration with anisotropic orthogonal Procrustes analysis
Author Affiliations +
Abstract
Real-time 3D ultrasound is an emerging imaging modality, offering full volumetric view of the anatomy without ionizing radiation. A spatial tracking system facilitates its integration with image-guided interventions, but such integration requires an accurate calibration between the spatial tracker and the ultrasound volume. In this paper, a rapid calibration technique for real-time 3D ultrasound is presented, comprising a plane-based calibration phantom, an algorithm for automatic fiducial extraction from ultrasound volumes, and a numerical solution for deriving calibration parameters involving anisotropic scaling. Using a magnetic tracking system and a commercial transesophageal echocardiogram real-time 3D ultrasound probe, this technique achieved a mean Target Registration Error of 2.9mm in a laboratory setting.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Elvis C. S. Chen, A. Jonathan McLeod, Uditha L. Jayarathne, Terry M. Peters, "Solving for free-hand and real-time 3D ultrasound calibration with anisotropic orthogonal Procrustes analysis", Proc. SPIE 9036, Medical Imaging 2014: Image-Guided Procedures, Robotic Interventions, and Modeling, 90361Z (12 March 2014); doi: 10.1117/12.2043080; https://doi.org/10.1117/12.2043080