Paper
1 May 2014 Sub-kHz traceable characterization of stroboscopic scanning white light interferometer
V. Heikkinen, I. Kassamakov, T. Paulin, A. Nolvi, J. Seppä, A. Lassila, E. Hæggström
Author Affiliations +
Abstract
Scanning white light interferometry (SWLI) is an established methodology for non-destructive testing of MEMS/NEMS. In contrast to monochromatic interference microcopy SWLI can unambiguously resolve surfaces featuring tall vertical steps. Oscillating samples can be imaged using a stroboscopic SWLI (SSWLI) equipped with a pulsed light source. To measure static samples the lateral and vertical scales of the SSWLI can be calibrated using transfer standards with calibrated dimensions such as line scales, 2D gratings, gauge blocks, and step height standards. However, traceable dynamic characterization of SSWLI requires a transfer standard (TS) providing repeatable traceable periodic movement. A TS based on a piezo-scanned flexure guided stage with capacitive feedback was designed and manufactured. The trajectories of the stage motion for different amplitude and frequency settings were characterized to have ~2 nm standard uncertainty. Characterization was made using a symmetric differential heterodyne laser interferometer (SDHLI). The TS was first used to characterize quasidynamic measurements across the vertical range of the SSWLI, 100 μm. Dynamic measurement properties of the SSWLI were then characterized using a sinusoidal vertical trajectory with 2 μm nominal amplitude and 50 Hz frequency. The motion amplitude of the TS, 2038 nm, measured with the SSWLI was 6 nm smaller than the amplitude measured with SDHLI. The repeatability of SSWLI expressed as experimental standard deviation of the mean was 8.8 nm. The maximum deviation in instantaneous displacement and oscillation velocity were 49 nm and 27 μm/s, respectively. A traceable method to characterize the capacity of the SSWLI to perform dynamic measurements at sub-kHz frequencies was demonstrated.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
V. Heikkinen, I. Kassamakov, T. Paulin, A. Nolvi, J. Seppä, A. Lassila, and E. Hæggström "Sub-kHz traceable characterization of stroboscopic scanning white light interferometer", Proc. SPIE 9132, Optical Micro- and Nanometrology V, 913218 (1 May 2014); https://doi.org/10.1117/12.2051622
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Signal to noise ratio

Interferometers

Light emitting diodes

Mirrors

Velocity measurements

Light sources

Back to Top