27 June 1988 Simulation Of A Single Photon Emission Computed Tomography Focusing Collimator System
John R. Baker, Thomas F. Budinger
Author Affiliations +
Abstract
A computer program to simulate the three dimensional impulse response of a convergent focus rectangular tube collimator system has been developed. The simulation is done by a deterministic algorithm that models geometric acceptance of photons, photon penetration of the septa, and single scatter interactions in the collimator walls. Output from the simulator is used as a basis set for the impulse response of a multiple detector SPECT system. The basis set is used by fitting procedures that estimate the impulse response of the Donner SPECT system (12 angles, 12x 128 detector positions per angle) from noisy measurements of a point source. These impulse response estimates are used to determine the weighting matrix of new emission reconstruction algorithms.
© (1988) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
John R. Baker and Thomas F. Budinger "Simulation Of A Single Photon Emission Computed Tomography Focusing Collimator System", Proc. SPIE 0914, Medical Imaging II, (27 June 1988); https://doi.org/10.1117/12.968677