In this paper, we report a field test of a microbend fiber sensor for simultaneous measurement of breathing rate, breathing pattern, Ballistocardiogram and heart rate during magnetic resonance imaging (MRI). Comparative experiments conducted between our sensor and commercial physiologic device on a healthy male subject showed an accuracy of +/-2bpm for simultaneous measurement of both breathing rate and heart rate. Our preliminary field test on simultaneous measurement of breathing rate and heart rate in a clinical trial conducted on 11 healthy subjects in the 3.0 Tesla MRI environment showed very good agreement compared with measurements obtained from conventional MRcompatible devices.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhihao Chen, Zhihao Chen, Doreen Lau, Doreen Lau, Ju Teng Teo, Ju Teng Teo, Soon Huat Ng, Soon Huat Ng, Xiufeng Yang, Xiufeng Yang, Pin Lin Kei, Pin Lin Kei, } "Field test of microbend fiber sensor for hospital use", Proc. SPIE 9157, 23rd International Conference on Optical Fibre Sensors, 91579A (2 June 2014); doi: 10.1117/12.2059417; https://doi.org/10.1117/12.2059417