5 September 2014 Spectroscopic response of Cd(Zn)Te radiation detectors with a Schottky diode
Author Affiliations +
Abstract
We investigated the spectroscopic properties of several Cd(Zn)Te detectors with a Schottky contact and simulated them via a computer code. The responses were determined of 0.5-mm-thick surface-barrier Ni/Cd(Zn)Te/Ni detectors to gamma-rays from reference sources of 241Am, 133Ba, 152Eu, 137Cs and 60Co. The best measured energy-resolution at 661.67 keV (137Cs) of these detectors under 800 V of displacement voltage was better than 1.5%. The detectors’ response functions, simulated with Geant4 toolkit, agreed satisfactorily with our experimental data.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Alexandr A. Zakharchenko, Alexandr A. Zakharchenko, Leonid N. Davydov, Leonid N. Davydov, Anna I. Skrypnyk, Anna I. Skrypnyk, Alexandr V. Rybka, Alexandr V. Rybka, Vladimir E. Kutny, Vladimir E. Kutny, Manap A. Khazhmuradov, Manap A. Khazhmuradov, Petro M. Fochuk, Petro M. Fochuk, Valery M. Sklyarchuk, Valery M. Sklyarchuk, Aleksey E. Bolotnikov, Aleksey E. Bolotnikov, Ralph B. James, Ralph B. James, } "Spectroscopic response of Cd(Zn)Te radiation detectors with a Schottky diode", Proc. SPIE 9213, Hard X-Ray, Gamma-Ray, and Neutron Detector Physics XVI, 92131C (5 September 2014); doi: 10.1117/12.2060403; https://doi.org/10.1117/12.2060403