Paper
25 November 2014 Improvement of the sensitivity of a stationary Raman gas analyzer
Author Affiliations +
Proceedings Volume 9292, 20th International Symposium on Atmospheric and Ocean Optics: Atmospheric Physics; 929229 (2014) https://doi.org/10.1117/12.2075413
Event: 20th International Symposium on Atmospheric and Ocean Optics: Atmospheric Physics, 2014, Novosibirsk, Russian Federation
Abstract
A prototype of a stationary Raman gas analyzer with the improved sensitivity is described. The improvement is provided by using both a device for compressing analyzed gas media and a specialized effective spectral device as a part of the gas analyzer. The experimental testing of the modified Raman gas analyzer was performed in the probing of atmospheric air that confirmed the extreme sensitivity of the prototype was equal to ~ 1 ppm.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
I. I. Matrosov, D. V. Petrov, and A. A. Tikhomirov "Improvement of the sensitivity of a stationary Raman gas analyzer", Proc. SPIE 9292, 20th International Symposium on Atmospheric and Ocean Optics: Atmospheric Physics, 929229 (25 November 2014); https://doi.org/10.1117/12.2075413
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Raman spectroscopy

Prototyping

Cameras

Atmospheric sensing

Atmospheric monitoring

Charge-coupled devices

Oxygen

Back to Top