28 April 2015 Preliminary results of non-contact THz imaging of cornea
Author Affiliations +
Abstract
This paper presents a novel THz optical design that allows the acquisition of THz reflectivity maps of in vivo cornea without the need for a field flattening window and preliminary imaging results of in vivo rabbit cornea. The system’s intended use is to sense small changes in corneal tissue water content (CTWC) that can be precursors for a host of diseases and pathologies. Unique beam optics allows the scanning of a curved surface at normal incidence while keeping the source detector and target stationary. Basic system design principles are discussed and image sets of spherical calibration targets and corneal phantom models are presented. The presented design will enable, for the first time, non-contact THz imaging of animal and human cornea.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Shijun Sung, James Garritano, Neha Bajwa, Sophie Deng, Jean-Pierre Hubschman, Warren S. Grundfest, Zachary D. Taylor, "Preliminary results of non-contact THz imaging of cornea", Proc. SPIE 9362, Terahertz, RF, Millimeter, and Submillimeter-Wave Technology and Applications VIII, 93620C (28 April 2015); doi: 10.1117/12.2086866; https://doi.org/10.1117/12.2086866
PROCEEDINGS
6 PAGES


SHARE
Back to Top