6 March 2015 Calibration of rotary table by whole combination measuring method
Author Affiliations +
Proceedings Volume 9446, Ninth International Symposium on Precision Engineering Measurement and Instrumentation; 94464E (2015) https://doi.org/10.1117/12.2181963
Event: International Symposium on Precision Engineering Measurement and Instrumentation, 2014, Changsha/Zhangjiajie, China
Abstract
Whole combination measuring method was used to calibrate a rotary table with a system error of 1", together with polygon; the inherent deviation of an angle in the polygon was then extracted from the results. The final PV value of angular measurement precision was ±0.5", which could be extended to the high-precision angular measurement.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hua Ai, Hua Ai, Yanbo Cao, Yanbo Cao, } "Calibration of rotary table by whole combination measuring method", Proc. SPIE 9446, Ninth International Symposium on Precision Engineering Measurement and Instrumentation, 94464E (6 March 2015); doi: 10.1117/12.2181963; https://doi.org/10.1117/12.2181963