Abstract
The possibilities of using gradient index (GRIN) lenses as the objective and relay lens in rigid endoscopic systems working in visible spectrum are investigated. The final aim of the project is to obtain a system providing diffraction limited image quality, which could be used for examining the gastrointestinal tract. This paper describes steps of the design of the system with working distance equal to 125 mm and angular field of view 25°. The diameter of the GRIN lenses is equal to 2,2 mm. Influence of the gradient of refractive index on the image quality is shown. Current results are analysed and future steps are described.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Grzegorz Fluder, Grzegorz Fluder, Anna Voznesenskaya, Anna Voznesenskaya, Alexey Bakholdin, Alexey Bakholdin, } "Design of rigid GRIN-endoscope with sapphire window and improved image quality ", Proc. SPIE 9529, Optical Methods for Inspection, Characterization, and Imaging of Biomaterials II, 95291N (22 June 2015); doi: 10.1117/12.2184843; https://doi.org/10.1117/12.2184843