31 August 2015 Digitizer architecture analysis for target diagnostics on the National Ignition Facility
Author Affiliations +
This paper covers a systems engineering analysis of existing scope-based Target Diagnostics (TD) on the National Ignition Facility (NIF) at Lawrence Livermore National Laboratory (LLNL), for the purpose of selecting a standard digitizer architecture future diagnostics. Key performance criteria and a summary of test results are presented.

Currently of the 60+ Target Diagnostics, at least fifteen use a type of high speed electrical signal data read-out device leading to over 200 digitization channels spread over six types of CRT and digital oscilloscopes, each with multiple models and versions. The proposed standard architecture discussed in this paper allows the NIF to efficiently and reliably operate digitizers that meet the required performance metrics for the lifetime of the NIF.

The systems engineering analysis identifies key stakeholders for multiple subsets of scope-based diagnostics including but not limited to the nToFs (neutron Time of Flight), DANTE a broadband, time-resolved x-ray spectrometer, SPBT (South Pole Bang Time), GRH (Gamma Reaction History), and FFLEX (Filter Fluorescer Diagnostic). From these stakeholders, key performance metrics are derived and feed into test and evaluation criteria for different digitizers and architectures.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
A. C. Carpenter, T. J. Clancy, B. Beeman, P. Bell, "Digitizer architecture analysis for target diagnostics on the National Ignition Facility", Proc. SPIE 9591, Target Diagnostics Physics and Engineering for Inertial Confinement Fusion IV, 95910B (31 August 2015); doi: 10.1117/12.2186102; https://doi.org/10.1117/12.2186102


Calibrating image plate sensitivity in the 700 to 5000 eV...
Proceedings of SPIE (September 26 2013)
Special tasks on FXR in detonics
Proceedings of SPIE (March 17 2005)
Our Concept Of Quality Control
Proceedings of SPIE (March 25 1976)
Calibrating of x ray detectors in the 8 to 111...
Proceedings of SPIE (October 19 2012)

Back to Top