You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
27 August 2015Dual-energy X-ray computed tomography system using a cadmium telluride detector and its application to gadolinium imaging
To obtain two tomograms with two different photon energy ranges simultaneously, we have performed dual-energy Xray photon counting using a cadmium telluride (CdTe) detector, two comparators, two frequency-voltage converters (FVCs), and an analog digital converter (ADC). X-ray photons are detected using the CdTe detector with an energy resolution of 1% at 122 keV, and the event pulses from a shaping amplifier are sent to two comparators simultaneously to regulate two thresholds of photon energy. The logical pulses from a comparator are sent to an FVC consisting of two integrators, a microcomputer, and a voltage-voltage amplifier. The smoothed outputs from the two FVCs are input to the ADC to carry out dual-energy imaging. To observe contrast variations with changes in threshold energy, we performed energy-dispersive computed tomography utilizing the dual-energy photon counting at a tube voltage of 100 kV and a current of 8.7 μA. Two tomograms were obtained simultaneously at two energy ranges of 34.0-50.2 keV and 50.2-100 keV. The photon-count subtraction was carried out using a computer program. The maximum count rate was 5.4 kilocounts per second with energies of 10.0-100 keV, and the exposure time for tomography was 10 min.
The alert did not successfully save. Please try again later.
Yasuhiro Miura, Eiichi Sato, Yasuyuki Oda, Satoshi Yamaguchi, Osahiko Hagiwara, Hiroshi Matsukiyo, Manabu Watanabe, Shinya Kusachi, "Dual-energy X-ray computed tomography system using a cadmium telluride detector and its application to gadolinium imaging," Proc. SPIE 9594, Medical Applications of Radiation Detectors V, 959402 (27 August 2015); https://doi.org/10.1117/12.2189262