23 December 1976 Acceptance Testing Experience With Fourteen New Installations
Phillip L Rauch, Roger W. Block
Author Affiliations +
Proceedings Volume 0096, Application of Optical Instrumentation in Medicine V; (1976) https://doi.org/10.1117/12.965385
Event: Application of Optical Instrumentation in Medicine V, 1976, Washington, D.C., United States
Abstract
One of the primary objectives of acceptance testing of a new radiological installation is the establishment of technique calibration of the x-ray generator. While we enthusiastically endorse the use of radiation penetrameters such as the Wisconsin Quality Control Cassette or a square root of 2 aluminum step wedge to assess the time dependent performance of an x-ray generator which is in clinical use, such tools are inappro priate for establishing calibration for a new facility. One must ascertain the behavior of the kilovoltage applied to the x-ray tube as a function of various conditions of loading and exposure control. That this is imperative should be made clear by the examples presented throughout this paper. The fourteen installations encompass five generator models produced by four manufacturers. Four of the generators are three phase twelve pulse. The fifth is further regulated to produce a nearly D.C. waveform. Three of these generators are furnished with mechanical type contacting while the other two are provided with silicon controlled rectifier exposure switching. The technique for observation of the performance of the kilovoltage waveform involves the introduction of a voltage divider or bleeder tank into the high tension secondary. It should be noted that all of the manufacturers who have installed equipment at Henry Ford Hospital utilize this technique. It has been our experience that kilovoltage calibration can be established to within 5%, and has often been obtained to better than 3% over the useful range of KV's and corresponding tube loads.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Phillip L Rauch and Roger W. Block "Acceptance Testing Experience With Fourteen New Installations", Proc. SPIE 0096, Application of Optical Instrumentation in Medicine V, (23 December 1976); https://doi.org/10.1117/12.965385
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Manufacturing

X-rays

Medicine

Optical instrument design

Chest imaging

Capacitors

RELATED CONTENT


Back to Top