23 December 1976 Development of a Phantom for Evaluation and Assurance of Image Quality in CT Scanning
David J. Goodenough, Kenneth E. Weaver, David O. Davis
Author Affiliations +
Proceedings Volume 0096, Application of Optical Instrumentation in Medicine V; (1976) https://doi.org/10.1117/12.965418
Event: Application of Optical Instrumentation in Medicine V, 1976, Washington, D.C., United States
Abstract
The development and use of a phantom for evaluation, comparison, and quality assurance of CT scanners will be discussed. Examples of measurements on seven CT scanners using early prototypes of the phantom will be presented along with measurements on several scanners using the final phantom configuration. The phantom contains four modular sections which are removable to allow for future fabrication and replacement of individual sections for specialized applications. Section I is used to measure contrast sensitivity and scan slice geometry of the system. Section II is used to measure the sensitometric response of the system. Section III is used to determine the spatial resolution of the system at various contrast levels. Section IV is used to determine the noise, spatial uniformity, alignment, and MTF of the system. In addition, it contains a part with fittings where items may be placed such as in vitro samples, dosimeters, or a motion phantom.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
David J. Goodenough, Kenneth E. Weaver, and David O. Davis "Development of a Phantom for Evaluation and Assurance of Image Quality in CT Scanning", Proc. SPIE 0096, Application of Optical Instrumentation in Medicine V, (23 December 1976); https://doi.org/10.1117/12.965418
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Computed tomography

Electromagnetic coupling

Signal detection

Signal attenuation

Scanners

Spatial resolution

Modulation transfer functions

Back to Top