7 August 2015 Spectral calibration for infrared hyperspectral imaging Fourier transform spectrometer based on absorption peaks
Author Affiliations +
Abstract
A new calibration method for infrared hyperspectral imaging Fourier transform spectrometer is presented. Two kinds of common materials as Polypropylene (PP) and Polyethylene Terephthalate (PET) films which have special absorption peaks in the infrared band were used in the calibration experiment. As the wavelengths at the sharp absorption peaks of the films are known, an infrared imaging spectrometer can be calibrated on spectra with two or three peaks. With high precision and stability, this method simplifies the calibration work. It is especially appropriate for the measuring condition with a lack of calibration equipment or with inconvenience to calibrate the multiple light sources outdoors.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
YaSheng Li, YaSheng Li, Yan Chen, Yan Chen, Ningfang Liao, Ningfang Liao, Hang Lyu, Hang Lyu, Shufang He, Shufang He, Lifang Wan, Lifang Wan, } "Spectral calibration for infrared hyperspectral imaging Fourier transform spectrometer based on absorption peaks", Proc. SPIE 9623, 2015 International Conference on Optical Instruments and Technology: Optoelectronic Measurement Technology and Systems, 96230J (7 August 2015); doi: 10.1117/12.2195687; https://doi.org/10.1117/12.2195687