12 October 2015 Creation and validation of Spectralon PTFE BRDF targets and standards
Author Affiliations +
Abstract
Sintered PTFE is an extremely stable, near-perfect Lambertian reflecting diffuser and calibration standard material that has been used by national labs, space, aerospace and commercial sectors for over two decades. New uncertainty targets of 2% on-orbit absolute validation in the Earth Observing Systems community have challenged the industry to improve is characterization and knowledge of almost every aspect of radiometric performance (space and ground). Assuming “near perfect” reflectance for angular dependent measurements is no longer going to suffice for many program needs. The total hemispherical spectral reflectance provides a good mark of general performance; but, without the angular characterization of bidirectional reflectance distribution function (BRDF) measurements, critical data is missing from many applications and uncertainty budgets. Therefore, traceable BRDF measurement capability is needed to characterize sintered PTFE’s angular response and provide a full uncertainty profile to users. This paper presents preliminary comparison measurements of the BRDF of sintered PTFE from several laboratories to better quantify the BRDF of sintered PTFE, assess the BRDF measurement comparability between laboratories, and improve estimates of measurement uncertainties under laboratory conditions.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Christopher Durell, Dan Scharpf, Greg McKee, Michelle L'Heureux, Georgi Georgiev, Gael Obein, and Catherine Cooksey "Creation and validation of Spectralon PTFE BRDF targets and standards", Proc. SPIE 9639, Sensors, Systems, and Next-Generation Satellites XIX, 96391D (12 October 2015); https://doi.org/10.1117/12.2195503
PROCEEDINGS
12 PAGES


SHARE
Advertisement
Advertisement
RELATED CONTENT

Ultraviolet BRDFs of selected materials
Proceedings of SPIE (November 11 1996)
Tracking TRMM/VIRS on-orbit calibration with MODIS
Proceedings of SPIE (September 27 2006)
The effect of speckle on BRDF measurements
Proceedings of SPIE (August 22 2005)

Back to Top