Paper
13 October 2015 Results of implementation of the dynamic laser goniometer for non-contact measurement of angular movement
Author Affiliations +
Abstract
The report presents results of implementation of the dynamic laser goniometer in the mode of non-contact measurements of an object’s angular position. One of obtained results is connected with determination of the time dependence of the scanning mirror angular position. Another kind of implementation result is determination of parameters of a test table oscillatory movement. The obtained results shows that the use of the LG makes it possible to calibrate various kinds of test-beds making angular oscillations or angular movement of some other law.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
E. Bokhman, M. Burnashev, Yu. Filatov, and P. Pavlov "Results of implementation of the dynamic laser goniometer for non-contact measurement of angular movement", Proc. SPIE 9649, Electro-Optical Remote Sensing, Photonic Technologies, and Applications IX, 96490N (13 October 2015); https://doi.org/10.1117/12.2194788