24 October 2005 An alternative method for evaluating the coherence length of a laser diode
Po-Jen Hsieh, Zhi-Cheng Jian, Der-Chin Su
Author Affiliations +
Proceedings Volume 9664, Ninth International Topical Meeting on Education and Training in Optics and Photonics; 96642N (2005) https://doi.org/10.1117/12.2207729
Event: Ninth International Topical Meeting on Education and Training in Optics and Photonics, 2005, Marseille, France
Abstract
Let a light beam coming from a diode pass through a polarizer and an electro-optical modulator driven a saw tooth voltage signal, then a slight frequency difference between its two linearly orthogonal polarized components occurs. The light beam enters a modified Michelson interferometer, and the interference signal is sent to an oscilloscope. Because the interference signal is a periodic function of time, it is easier than the conventional one to evaluate the contrast. When the contrast is zero, the associated optical path difference is its coherence length.

Summary

The schematic diagram of this method is shown in Fig. 1. The light beam coming from a laser diode LD passing through a polarizer P and an electro-optic modulator EO incidents on a beam splitter BS. The transmitted light enters a modified Michelson interferometer[1] and is divided into two parts by a polarization beam splitter PBS: the reflected s-polarization light and the transmitted p-polarization light. The former is normally reflected by a mirror M1, and then is reflected by the PBS and the BS again. And it passes through an analyzer AN and is detected by a photo detector PD. It acts the reference light in this interferometer. The latter is normally reflected by another mirror M2 and returns along its original path. After being reflected by the BS, it passes through the AN and also enters the PD. It acts as the test light in the modified Michelson interferometer. If the amplitudes of the reference light and the test light are Er and Et, PD measures the interference intensity of Er and Et, that is, l= |Er+Et|2.

Fig. 1

The schematic diagram of this method for measuring the coherence length of a laser diode. P: polarizer; EO: electro-optic modulator; BS: beam splitter; PBS: polarization beam splitter; M: mirror; AN: analyzer; PD: photo detector; OSC: oscilloscope.

00094_psisdg9664_96642N_page_3_1.jpg

For convenience, the +z axis is chosen along the propagation direction, the x-axis is along the horizontal direction and only the light of the wavelength λi is considered. et the transmission axis of the P be at 45˚ with respect o the x-axis, then the Jone vector[2] of the light incidenting on the BS can be written as

00094_psisdg9664_96642N_page_1_1.jpg

If the fast axis of the EO is along the x-axis, and an external saw tooth voltage signal with frequency f and amplitude V is applied to the EO, then the retardation produced by the EO can be expressed as[3,4]

00094_psisdg9664_96642N_page_1_2.jpg

where V is the associated half-wave voltage of the EO at λi. And if the transmission axis of the AN is 45˚ with respect to the x-axis, then we have

00094_psisdg9664_96642N_page_1_3.jpg

and

00094_psisdg9664_96642N_page_2_1.jpg

where ϕi is the phase difference being corresponding to the optical path difference between the test light and the reference light, i.e.,

00094_psisdg9664_96642N_page_2_2.jpg

Therefore, the interference intensity is given by

00094_psisdg9664_96642N_page_2_3.jpg

Because the light coming from a laser diode contains many continuous spectrums, the interference intensity measured by the PD can be expressed as

00094_psisdg9664_96642N_page_2_4.jpg

where n is an integer, and ai is the bias intensity of wavelength λi. From Eq. (7), it is obvious that we can obtain the summation signal of many cosine signals on the oscilloscope OSC. Then, moving M2 in steps until the contrast of the signal on the OSC is nearly equivalent to zero. At the time, the optical path difference between the test light and the reference light is the coherence length.

In order to show the validity of this method, a laser diode (HHL6720G) manufactured by Japan Hitachi Ltd. is tested. It operated at 25 °C with electric current 35 mA and its central wavelength is 670.57 nm. An EO modulator (4002) fabricated by New Focus, with half-wave voltage 220 V at 670.57 nm, is used in this test. A saw tooth signal, with frequency 2 kHz and amplitude

220 V, is applied to the EO modulator. Fig. 2 are the interference signal on the OSC as the optical path difference is about (a) zeromm, (b) 27mm, (c) 42mm, respectively. According Fig. 2, we obtain its coherence length is about 42mm. In addition, we also measure its coherence length with the conventional Michelson interferometer. We obtain the same result.

Fig. 2

The interference signals on the OSC as the optical path difference is about(a) 0 mm, (b) 27 mm, and (c) 42 mm, respectively.

00094_psisdg9664_96642N_page_3_2.jpg

The half-voltage of the EO is varied as the wavelength is varied[5]. In our test, the tested LD has narrow spectral bandwidth, the condition VVi exists in Eq. (7). So our measured result has the same result as the conventional method. If a light source with wide spectral bandwidth or several separate spectral lines[6], is tested, the half-voltage is different for different wavelength. Consequently, the measured result with this method will be smaller than that of the conventional method. We also measure two He-Ne lasers (GLG 5369 and GLG 5730) manufactured by Japan NEC Ltd. The half-voltage 146.4 V at 632.8 nm is applied to the EO. Their measured results are 256 mm and 378 mm (with this method), and 305 mm and 450 mm (with the conventional method), respectively. Therefore this method is suitable only for the light source with narrow spectral bandwidth. In addition, either mirror does not move as the contrast of the interference signal is monitored in the process of this method, so this method has some merits such as high stable signal, easy operation, and high accuracy.

Acknowledgement

This study was supported in part by the National Science Council, Taiwan, under contract NSC 93-2215-E-009-021.

References

[1] 

P. Hariharan, “Optical interferometery,” Academic Press, 22003). Google Scholar

[2] 

A. Yariv and P. Yeh, “Optical waves in crystals,” 5 John Wiley and Sons, Inc., N. Y., Chap (1984). Google Scholar

[3] 

D. C. Su, M. H. Chiu, andC. D. Chen, “Simple two-frequency laser,” Precision Engineering, 18 161 –163 https://doi.org/10.1016/0141-6359(96)80485-5 Google Scholar

[4] 

D. C. Su, M. H. Chiu, and C. D. Chen, “A heterodyne interferometer using an electro-optic modulator for measuring small displacements,” J. Optics, 27 19 –23 (1996). https://doi.org/10.1088/0150-536X/27/1/003 Google Scholar

[5] 

A. Yariv and P. Yeh, “Optical waves in crystals,” 7 John Wiley and Sons, Inc., N. Y., Chap (1984). Google Scholar

[6] 

O. Svelto and D. C. Hanna, “Principles of Lasers,” Plenum Press, N. Y. (1989). Google Scholar
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Po-Jen Hsieh, Zhi-Cheng Jian, and Der-Chin Su "An alternative method for evaluating the coherence length of a laser diode", Proc. SPIE 9664, Ninth International Topical Meeting on Education and Training in Optics and Photonics, 96642N (24 October 2005); https://doi.org/10.1117/12.2207729
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Semiconductor lasers

Michelson interferometers

Coherence (optics)

Modulators

Teeth

Beam splitters

Electrooptic modulators

Back to Top