You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
8 October 2015Focal length and radius of curvature measurement using wavefront difference method
A method for measuring the focal length of the lens and the radius of curvature of the spherical surface using wavefront difference method is proposed. Based on Fizeau interferometer, an experimental system for focal length measurements is set up to verify the principle. Based on the point diffraction interferometer, an experimental system for radius of curvature measurements is proposed to verify the proposed method. With the focal length testing system, both the positive and negative lens experimental results indicate that the measurement accuracy is less than 0.16%under normal experimental environment. With the radius of curvature testing system, the radius of curvature of spherical mirrors and the surface figure can be measured in a higher precision simultaneously. The experimental results indicate that the measurement accuracy is in the order of 10-4 .