7 March 2016 Diagnosis potential of near infrared Mueller Matrix imaging for colonic adenocarcinoma
Author Affiliations +
Abstract
Mueller matrix imaging along with polar decomposition method was employed for the colonic adenocarcinoma detection by polarized light in the near-infrared spectral range (700–1100 nm). A high-speed (<5s) Muller matrix imaging system with dual-rotating waveplates was developed. 16 (4 by 4) full Mueller matrices of the colonic tissues (i.e., normal and caner) were acquired. Polar decomposition was further implemented on the 16 images to derive the diattentuation, depolarization, and the retardance images. The decomposed images showed clear margin between the normal and adenocarcinomaous colon tissue samples. The work shows the potential of near-infrared Mueller matrix imaging for the early diagnosis and detection of malignant lesions in the colon.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jianfeng Wang, Jianfeng Wang, Wei Zheng, Wei Zheng, Kan Lin, Kan Lin, Zhiwei Huang, Zhiwei Huang, } "Diagnosis potential of near infrared Mueller Matrix imaging for colonic adenocarcinoma", Proc. SPIE 9698, Advanced Biomedical and Clinical Diagnostic and Surgical Guidance Systems XIV, 96980X (7 March 2016); doi: 10.1117/12.2212074; https://doi.org/10.1117/12.2212074
PROCEEDINGS
5 PAGES


SHARE
Back to Top