You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
18 March 2016Validation of MTF measurement for CBCT system using Monte Carlo simulations
To evaluate the spatial resolution performance of cone beam computed tomography (CBCT) system, accurate measurement of the modulation transfer function (MTF) is required. This accuracy depends on the MTF measurement method and CBCT reconstruction algorithms. In this work, the accuracy of MTF measurement of CBCT system using wire phantom is validated by Monte Carlo simulation. A Monte Carlo simulation software tool BEAMnrc/EGSnrc was employed to model X-ray radiation beams and transport. Tungsten wires were simulated with different diameters and radial distances from the axis of rotation. We adopted filtered back projection technique to reconstruct images from 360° acquisition. The MTFs for four reconstruction kernels were measured from corresponding reconstructed wire images, while the ram-lak kernel increased the MTF relative to the cosine, hamming and hann kernel. The results demonstrated that the MTF degraded radially from the axis of rotation. This study suggested that an increase in the MTF for the CBCT system is possible by optimizing scanning settings and reconstruction parameters.
The alert did not successfully save. Please try again later.
Ting Hao, Feng Gao, Huijuan Zhao, Zhongxing Zhou, "Validation of MTF measurement for CBCT system using Monte Carlo simulations," Proc. SPIE 9700, Design and Quality for Biomedical Technologies IX, 970017 (18 March 2016); https://doi.org/10.1117/12.2210746