10 March 2016 Cross-calibrating interferon-γ detection by using eletrochemical impedance spectroscopy and paraboloidal mirror enabled surface plasmon resonance interferometer
Author Affiliations +
Proceedings Volume 9701, Multimodal Biomedical Imaging XI; 97010U (2016) https://doi.org/10.1117/12.2211494
Event: SPIE BiOS, 2016, San Francisco, California, United States
Abstract
Tuberculosis is a highly contagious disease such that global latent patient can be as high as one third of the world population. Currently, latent tuberculosis was diagnosed by stimulating the T cells to produce the biomarker of tuberculosis, i.e., interferon-γ. In this paper, we developed a paraboloidal mirror enabled surface plasmon resonance (SPR) interferometer that has the potential to also integrate ellipsometry to analyze the antibody and antigen reactions. To examine the feasibility of developing a platform for cross calibrating the performance and detection limit of various bio-detection techniques, electrochemical impedance spectroscopy (EIS) method was also implemented onto a biochip that can be incorporated into this newly developed platform.

The microfluidic channel of the biochip was functionalized by coating the interferon-γ antibody so as to enhance the detection specificity. To facilitate the processing steps needed for using the biochip to detect various antigen of vastly different concentrations, a kinetic mount was also developed to guarantee the biochip re-positioning accuracy whenever the biochip was removed and placed back for another round of detection. With EIS being utilized, SPR was also adopted to observe the real-time signals on the computer in order to analyze the success of each biochip processing steps such as functionalization, wash, etc. Finally, the EIS results and the optical signals obtained from the newly developed optical detection platform was cross-calibrated. Preliminary experimental results demonstrate the accuracy and performance of SPR and EIS measurement done at the newly integrated platform.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Meng-Wei Liu, Hao-Jung Chang, Shu-sheng Lee, and Chih-Kung Lee "Cross-calibrating interferon-γ detection by using eletrochemical impedance spectroscopy and paraboloidal mirror enabled surface plasmon resonance interferometer", Proc. SPIE 9701, Multimodal Biomedical Imaging XI, 97010U (10 March 2016); https://doi.org/10.1117/12.2211494
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Interferometers

Surface plasmons

Dielectric spectroscopy

Kinematics

Mirrors

Coating

Signal detection

RELATED CONTENT

IASI instrument overview
Proceedings of SPIE (December 15 1995)
Rocket-Borne Cryogenic Michelson Interferometer
Proceedings of SPIE (October 29 1979)
Nulling at the Keck interferometer
Proceedings of SPIE (June 23 2006)

Back to Top