7 March 2016 Improved fiber probe for laser tissue ablation with integrated distributed temperature sensor
Author Affiliations +
Abstract
The paper presents a new all-fiber probe for laser induced thermal ablation of solid tumor cells that integrates a beam delivery fiber with nanostructured surface to shape the laser irradiation pattern and a chirped grating to allow real-time monitoring of the temperature profile. A theoretical model of the sensor to study the temperature profile recovery algorithm and experimental validations using phantoms are discussed.
Conference Presentation
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yu Liu, Yu Liu, Riccardo Gassino, Riccardo Gassino, Hao Yu, Hao Yu, Andrea Braglia, Andrea Braglia, Alberto Vallan, Alberto Vallan, Guido Perrone, Guido Perrone, Daniele Tosi, Daniele Tosi, } "Improved fiber probe for laser tissue ablation with integrated distributed temperature sensor", Proc. SPIE 9702, Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications XVI, 97020G (7 March 2016); doi: 10.1117/12.2211292; https://doi.org/10.1117/12.2211292
PROCEEDINGS
7 PAGES + PRESENTATION

SHARE
Back to Top