25 March 2016 Crystal timing offset calibration method for time of flight PET scanners
Jinghan Ye, Xiyun Song
Author Affiliations +
Abstract
In time-of-flight (TOF) positron emission tomography (PET), precise calibration of the timing offset of each crystal of a PET scanner is essential. Conventionally this calibration requires a specially designed tool just for this purpose. In this study a method that uses a planar source to measure the crystal timing offsets (CTO) is developed. The method uses list mode acquisitions of a planar source placed at multiple orientations inside the PET scanner field-of-view (FOV). The placement of the planar source in each acquisition is automatically figured out from the measured data, so that a fixture for exactly placing the source is not required. The expected coincidence time difference for each detected list mode event can be found from the planar source placement and the detector geometry. A deviation of the measured time difference from the expected one is due to CTO of the two crystals. The least squared solution of the CTO is found iteratively using the list mode events. The effectiveness of the crystal timing calibration method is evidenced using phantom images generated by placing back each list mode event into the image space with the timing offset applied to each event. The zigzagged outlines of the phantoms in the images become smooth after the crystal timing calibration is applied. In conclusion, a crystal timing calibration method is developed. The method uses multiple list mode acquisitions of a planar source to find the least squared solution of crystal timing offsets.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jinghan Ye and Xiyun Song "Crystal timing offset calibration method for time of flight PET scanners", Proc. SPIE 9783, Medical Imaging 2016: Physics of Medical Imaging, 97831M (25 March 2016); https://doi.org/10.1117/12.2217211
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Crystals

Calibration

Positron emission tomography

Scanners

Sensors

Data acquisition

Time metrology

RELATED CONTENT

A simple all-digital PET system
Proceedings of SPIE (March 15 2007)
Analysis system for PET detector
Proceedings of SPIE (July 04 2000)

Back to Top