10 January 2017 Erratum: Monte Carlo investigation of the dosimetric effect of the Autoscan ultrasound probe for guidance in radiotherapy
Author Affiliations +
Abstract
A revised version of this paper, published originally on 2016, was published on 6 January, 2017, replacing the original paper. To correct for a set-up error, the simulated treatment plan was re-run using CT images of a solid water phantom with/without the Autoscan US probe in contact with its surface. The revised paper is available at http://dx.doi.org/10.1117/12.2216653.

Changes to original text:

  • Section 2.2
  • 1. A solid water phantom is employed as opposed to a Rando phantom.

    2. Figure 1 is changed to illustrate new phantom.

  • Section 2.3
  • 1. New treatment isocentre.

    2. Figure 3 is changed to illustrate dose distribution in new phantom.

  • Section 3.2
  • 1. Text of first paragraph has been revised to reflect new results.

    2. Table 2 has been revised to reflect new results.

    3. Figure 6 has been revised to reflect new results.

  • Section 4
  • 1. Text of second paragraph has been revised to reflect new results.

    © (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
    Michael Martyn, Michael Martyn, Tuathan O'Shea, Tuathan O'Shea, Emma Harris, Emma Harris, Jeffrey C. Bamber, Jeffrey C. Bamber, Stephen Gilroy, Stephen Gilroy, Mark J. Foley, Mark J. Foley, } "Erratum: Monte Carlo investigation of the dosimetric effect of the Autoscan ultrasound probe for guidance in radiotherapy", Proc. SPIE 9790, Medical Imaging 2016: Ultrasonic Imaging and Tomography, 97902A (10 January 2017); doi: 10.1117/12.2271153; https://doi.org/10.1117/12.2271153