12 May 2016 High precision, medium flux rate CZT spectroscopy for coherent scatter imaging
Author Affiliations +
Abstract
CZT detectors are primary candidates for many next-generation X-ray imaging systems. These detectors are typically operated in either a high precision, low flux spectroscopy mode or a low precision, high flux photon counting mode. We demonstrate a new detector configuration that enables operation in a high precision, medium flux spectroscopy mode, which opens the potential for a variety of new applications in medical imaging, non-destructive testing and baggage scanning. In particular, we describe the requirements of a coded aperture coherent scattering X-ray system that can perform fast imaging with accurate material discrimination.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Joel A. Greenberg, Mehadi Hassan, David J. Brady, and Kris Iniewski "High precision, medium flux rate CZT spectroscopy for coherent scatter imaging", Proc. SPIE 9847, Anomaly Detection and Imaging with X-Rays (ADIX), 98470E (12 May 2016); https://doi.org/10.1117/12.2221913