18 July 2016 Alignment and calibration of the ICON-FUV instrument: development of a vacuum UV facility
Author Affiliations +
Abstract
The optical calibration of the ICON-FUV instrument requires designing specific ground support equipment (GSE). The ICON-FUV instrument is a spectrographic imager that operates on two specific wavelengths in the UV (135.6 nm and 157 nm). All the operations have to be performed under vacuum UV light. The optical setup is based on a VUV monochromator coupled with a collimator that illuminates the FUV entrance slit. The instrument is placed on a manipulator providing fields pointing. Image quality and spectral properties can be then characterized for each field. OGSE, MGSE, optical calibration plan and vacuum alignment of the instrument are described.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jerome Loicq, Jerome Loicq, Pascal Blain, Pascal Blain, Christian Kintziger, Christian Kintziger, Richard Desselle, Richard Desselle, Yvan Stockman, Yvan Stockman, Isabelle Domken, Isabelle Domken, Etienne Renotte, Etienne Renotte, Lionel Clermont, Lionel Clermont, Kodi Rider, Kodi Rider, Cathy Chou, Cathy Chou, Claire Poppett, Claire Poppett, Harald U. Frey, Harald U. Frey, Stephen B. Mende, Stephen B. Mende, "Alignment and calibration of the ICON-FUV instrument: development of a vacuum UV facility", Proc. SPIE 9905, Space Telescopes and Instrumentation 2016: Ultraviolet to Gamma Ray, 99052W (18 July 2016); doi: 10.1117/12.2233192; https://doi.org/10.1117/12.2233192
PROCEEDINGS
15 PAGES


SHARE
Back to Top