26 July 2016 Ground calibration of the Astro-H (Hitomi) soft x-ray spectrometer
M. E. Eckart, J. S. Adams, K. R. Boyce, G. V. Brown, M. P. Chiao, R. Fujimoto, D. Haas, J. W. den Herder, Y. Ishisaki, R. L. Kelley, C. A. Kilbourne, M. A. Leutenegger, D. McCammon, K. Mitsuda, F. S. Porter, K. Sato, M. Sawada, H. Seta, G. A. Sneiderman, A. E. Szymkowiak, Y. Takei, M. Tashiro, M. Tsujimoto, C. P. de Vries, T. Watanabe, S. Yamada, N. Y. Yamasaki
Author Affiliations +
Abstract
The Astro-H (Hitomi) Soft X-ray Spectrometer (SXS) was a pioneering imaging x-ray spectrometer with 5 eV energy resolution at 6 keV. The instrument used a microcalorimeter array at the focus of a high-throughput soft x-ray telescope to enable high-resolution non-dispersive spectroscopy in the soft x-ray waveband (0:3-12 keV). We present the suite of ground calibration measurements acquired from 2012-2015, including characterization of the detector system, anti-coincidence detector, optical blocking filters, and filter-wheel filters. The calibration of the 36-pixel silicon thermistor microcalorimeter array includes parameterizations of the energy gain scale and line spread function for each event grade over a range of instrument operating conditions, as well as quantum efficiency measurements. The x-ray transmission of the set of five Al/polyimide thin-film optical blocking filters mounted inside the SXS dewar has been modeled based on measurements at synchrotron beamlines, including with high spectral resolution at the C, N, O, and Al K-edges. In addition, we present the x-ray transmission of the dewar gate valve and of the filters mounted on the SXS filter wheel (external to the dewar), including beryllium, polyimide, and neutral density filters.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
M. E. Eckart, J. S. Adams, K. R. Boyce, G. V. Brown, M. P. Chiao, R. Fujimoto, D. Haas, J. W. den Herder, Y. Ishisaki, R. L. Kelley, C. A. Kilbourne, M. A. Leutenegger, D. McCammon, K. Mitsuda, F. S. Porter, K. Sato, M. Sawada, H. Seta, G. A. Sneiderman, A. E. Szymkowiak, Y. Takei, M. Tashiro, M. Tsujimoto, C. P. de Vries, T. Watanabe, S. Yamada, and N. Y. Yamasaki "Ground calibration of the Astro-H (Hitomi) soft x-ray spectrometer", Proc. SPIE 9905, Space Telescopes and Instrumentation 2016: Ultraviolet to Gamma Ray, 99053W (26 July 2016); https://doi.org/10.1117/12.2233053
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical filters

Silicon

X-rays

X-rays

Sensors

Thin films

Interfaces

RELATED CONTENT

Status of the EPIC thin and medium filters on board...
Proceedings of SPIE (September 26 2013)
Status of the wide field imager instrument for Athena
Proceedings of SPIE (September 09 2019)
On orbit ScaRaB flight model 2 calibration procedure
Proceedings of SPIE (September 18 1997)
Thin metal film thermal microsensors
Proceedings of SPIE (September 26 1995)

Back to Top