22 July 2016 Low cost photonic comb for sub-m/s wavelength calibration
Author Affiliations +
Abstract
A fundamental limitation of precision radial velocity measurements is the accuracy and stability of the calibration source. Here we present a low-cost alternative to more complex laser metrology based systems that utilises a single-mode fibre Fabry-Perot etalon. There are three key elements on this photonic comb: i) an optical fibre etalon with thermo-electric coolers; ii) a Rubidium Saturation Absorption Spectroscopy (SAS) setup; and iii) an optical fibre switch system for simultaneous laser locking of the etalon. We simultaneously measure the Rubidium D2 transitions around 780.2 nm and the closest etalon line. A PID loop controls the etalon temperate to maintain the position of its peak with an RMS error of <10cm/s for 10 minute integration intervals in continous operation. The optical fibre switch system allows for a time multiplexed coupling of the etalon to a spectrograph and SAS system.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Christopher H. Betters, Christopher H. Betters, Maxime Hermouet, Maxime Hermouet, Thomas Blanc, Thomas Blanc, James I. Colless, James I. Colless, Joss Bland-Hawthorn, Joss Bland-Hawthorn, Janez Kos, Janez Kos, Sergio Leon-Saval, Sergio Leon-Saval, } "Low cost photonic comb for sub-m/s wavelength calibration", Proc. SPIE 9912, Advances in Optical and Mechanical Technologies for Telescopes and Instrumentation II, 99126Y (22 July 2016); doi: 10.1117/12.2233678; https://doi.org/10.1117/12.2233678