19 September 2016 Data processing for a multi-slit LWIR HSI spectrometer
Author Affiliations +
Abstract
Preliminary testing of a three-slit prism-based spectrometer is presented to test means of exploiting data from a multi-slit spectrometer as well as some potential ways of dealing with complications that arise when using multiple slits. When using a multiple slit spectrometer to boost SNR there are two primary concerns: first, the spectral axis of each slit must be nearly identical to effective average and second, the image from each slit must be well-registered. Based on some of these complications it seems, given the current technology, the best operational mode is to use the sensor to increase area coverage.
Conference Presentation
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jacob A. Martin, Jacob A. Martin, Joseph Meola, Joseph Meola, } "Data processing for a multi-slit LWIR HSI spectrometer", Proc. SPIE 9976, Imaging Spectrometry XXI, 997607 (19 September 2016); doi: 10.1117/12.2239123; https://doi.org/10.1117/12.2239123