1 November 2005 Optoacoustic real-time dosimetry for selective retina treatment
Georg Schuele, Hanno Elsner, Carsten Framme, Johann Roider M.D., Reginald Birngruber, Ralf Brinkmann
Author Affiliations +
Abstract
The selective retina treatment (SRT) targets retinal diseases associated with disorders in the retinal pigment epithelium (RPE). Due to the ophthalmoscopic invisibility of the laser-induced RPE effects, we investigate a noninvasive optoacoustic real-time dosimetry system. In vitro porcine RPE is irradiated with a Nd:YLF laser (527 nm, 1.7-µs pulse duration, 5 to 40 µJ, 30 pulses, 100-Hz repetition rate). Generated acoustic transients are measured with a piezoelectric transducer. During 27 patient treatments, the acoustic transients are measured with a transducer embedded in an ophthalmic contact lens. After treatment, RPE damage is visualized by fluorescein angiographic leakage. Below the RPE damage threshold, the optoacoustic transients show no pulse-to-pulse fluctuations within a laser pulse train. Above threshold, fluctuations of the individual transients among each other are observed. If optoacoustic pulse-to-pulse fluctuations are present, RPE leakage is observed in fluorescein angiography. In 96% of the irradiated areas, RPE leakage correlated with the optoacoustic defined threshold value. A noninvasive optoacoustic real-time dosimetry for SRT is developed and proved in vitro and during patient treatment. It detects the ophthalmoscopically invisible laser-induced damage of RPE cells and overcomes practical limitations of SRT for use in private practice.

1.

Introduction

The first use of lasers in medicine was for the treatment of retinal diseases. Due to the optical transparency of the eye, laser energy can be effectively delivered to the retina. Continuous wave (cw) photocoagulation became the gold standard treatment for several retinal diseases.1, 2, 3 Photocoagulation of the retina with a cw laser ( 50to300-ms exposure time) results in visible laser burns of the retina, causing immediate functional loss in the treated area, a so-called scotoma. Novel retinal treatment modalities try to induce a specific therapeutic effect, but avoid damage of the photoreceptors and neural retina by selective targeting of different fundus structures. The mechanisms behind these techniques range from a photochemical effect in photodynamic therapy (PDT),4 pure thermal in transpupillary thermotherapy (TTT),5 to thermomechanical in selective retina treatment (SRT).6, 7, 8 Due to the high inter- and also intra-individual variability of the optical transmission of the eye, blood perfusion of the choroidal vessels, and light absorption within the fundus, a real-time dosimetry method is preferable to ensure the selectivity and to monitor the therapeutic treatment progress.

In SRT, a train of microsecond laser pulses is applied to the retina.6 Up to 60% of the incident light is absorbed by the melanosomes inside the retinal pigment epithelium (RPE) cells.9 Since the exposure time is short compared to the thermal relaxation of the several micrometer thick RPE layer, the induced heat is confined to the RPE cell layer. This results in selective heating of the RPE cell layer. Calculations show that the induced temperatures at the adjacent photoreceptors are substantially lower than in the RPE.10 It has been shown that applying a train of microsecond (μs) laser pulses destroys the RPE selectively, thus maintaining full vision in the treated area.8 This technique allows treatment in proximity of the fovea.

SRT targets different diseases that are associated with decreased metabolic activity of the RPE. The aim of SRT is the rejuvenation of the RPE by inducing a RPE healing process. So far, more than 200 patients with diabetic maculopathy, central serous retinopathy, or drusen maculopathy were treated in a preclinical study. The rationale for treating these diseases with SRT is described next.

Diabetic maculopathy (DMP) is a disease of the microvascular system caused by high sugar levels over a long period of time. The best predictor of diabetic retinopathy is the duration of the disease.11 Patients who have had diabetic type 1 for more than 20years show an incidence of 95% for DMP. In diabetic maculopathy, the altered microvascular system leads to localized or diffuse fluid accumulation within the retina, forming cystoid pockets of fluid in the outer plexiform layer in late stages of the disease. This retinal thickening leads to profound functional loss. The rational for treating DMP with lasers is to alter the RPE to increase metabolic activity and to thereby reduce retinal edema.12 This would stabilize or even improve visual function. However, conventional argon laser treatment leads to irreversible photoreceptor damage and to microscotoma. Especially in cases with diabetic pathology close to the fovea, conventional argon laser treatment is no option. SRT treats the RPE while sparing retinal structures. Pathology close to the fovea can be tackled.

Central serous retinopathy (CSR) is a common retinal disorder characterized by an idiopathic serous neural retinal detachment in the macular region. Typically, CSR affects men aged 20to50years . The typical CSR case shows a small leak on the level of the RPE, which can be identified easily by fluorescein angiography. Through this leak, serous fluid from the choroid detaches the retina. If the fovea is affected, the patient realizes functional loss, which is reversible if the duration of the retinal detachment is short. By applying only several laser spots to the RPE leakage site, the area is treated. However, 75% of the leaking sites are located within one disk area close to the fovea, so a conventional argon laser has substantial risk of damaging retinal tissue close to the fovea. SRT selectively closes the RPE leak and spares retinal tissue, and is therefore the only therapeutical option for treating the case with RPE leaks close to the fovea.

Age-related macular degeneration (AMD) is the leading cause of central visual loss among individuals 65years of age and older in developed countries.13, 14, 15, 16 The disease primarily affects the choriocapillaris, Bruch’s membrane, and RPE. The average age at onset of visual loss is about 75years . In early stages, metabolic deposits (drusen) are localized between the RPE cells and Bruchs membrane, and are precursors of advanced AMD, either dry geographic or neovascular AMD. Drusen are a sign of collapsing metabolic transport from the RPE cells to the choroid occurring with age. The rational of treating drusen with SRT is to selectively damage RPE cells in the area of drusen to improve metabolic activity. This can reduce the amount and number of drusen material, and could positively influence the natural course of the disease and prevent development of late stage AMD.

In SRT, the selective RPE damage is ophthalmoscopically invisible but can be visualized invasively after treatment by fluorescein or indio-cyanine green (ICG) angiography.6, 7, 8 This implies several limitations of this novel treatment for common use. First, treatment is only possible if also an angiography device is accessible. Second, the success of the laser treatment is only visible after an angiography. Third, angiography generally is an invasive technique (intravenous dye injection) with the risk of allergic reactions, anaphylactic shock, or death.

The formation of microbubbles around the strong absorbing melanosomes inside the RPE has proved to be the damage mechanism during irradiation of the RPE with μs laser pulses.10, 17 In this case, mechanical effects such as cell disruption from intracellular microbubbles most likely induce RPE cell damage. For laser pulses shorter than several nanoseconds, cell damage can also be induced by microbubbles or by stress waves.18, 19 For longer laser pulses (> 500μs) , the RPE cells are assumed to be thermally denaturated.20, 21

If energy is absorbed and converted to heat, the thermoelastic expansion of the absorbing medium will generate an optoacoustic (OA) transient.22, 23 OA techniques have been used in ophthalmology to monitor the laser-induced retinal temperature increase in vitro during continuous wave (cw) photocoagulation and during SRT patient treatment.24 Also, OA imaging of the ciliary body and OA monitoring of transscleral cyclophotocoagulation is possible in vitro.25 In another in vitro study, OA techniques are used for temperature mapping of tissue during laser-induced thermotherapy (LITT).26, 27, 28

During irradiation of RPE with μs laser pulses, an OA transient will be produced.24 Due to the formation and collapse of microbubbles around the melanosomes during a successful SRT, additional OA bubble transients will be emitted.29 This is analogous to the emission of acoustic transients during formation and collapse of cavitation bubbles.30

The objective of this study was to develop and validate a real-time OA dosimetry system for SRT. This system should reliably detect the laser-induced microbubble formation around the melanosomes during a successful treatment of irradiation.

2.

Material and Methods

2.1.

In Vitro Experiments

2.1.1.

Setup for in vitro experiments

An intracavity frequency-doubled, pulse-stretched Nd:YLF laser ( 527nm , 1.7-μs pulse duration, 100-Hz repetition rate) was used as an irradiation source. A more detailed description of the laser itself can be found elsewhere.10 The fiber tip was imaged with an ophthalmic laser slit lamp (Zeiss, 30 SL/L) to the sample, which was fixed in a water-filled cuvette (Fig. 1 ). The beam diameter and profile in the sample plane was measured with a laser beam analyzer (LBA-300PC, Spiricon Incorporated, Logan, UT). The beam profile was a tophat with an intensity variation of 15% due to speckle formation. The beam diameter was 160μm in air. Acoustic transients were received with a commercial available ultrasonic broadband transducer (Valpey-Fisher, VP-1093, 0to10MHz ), preamplified (Panametrics, Model 5676, 54-dB amplification), and recorded by a transient recorder (TEK/Sony, RTD710). The distance between sample and transducer was approximately 2mm . A hydrophone (Ceram number 118) generating 15mVbar was used to calibrate the broadband ultrasonic transducer, yielding a sensitivity of 1.05Vbar at 1MHz . For calibration, both detectors were placed in a water bath. They were adjusted to the same angle and the same distance to a black absorber sample. Laser pulses with 150ns in duration and 20to100-μJ pulse energy were applied to the sample, and the amplitude of the generated thermoelastic transient was measured simultaneously with an oscilloscope (TEK, TDS220).

Fig. 1

Setup for optoacoustic measurements during irradiation of porcine RPE with μs -laser pulses.

064022_1_013506jbo1.jpg

2.1.2.

In vitro retinal pigment epithelium sample preparation, vitality staining, and laser parameters

The experiments were performed on six freshly enucleated porcine eyes from the slaughterhouse. After equatorial dissection, the vitreous, neural, and neurosensory tissue, including the photoreceptors, were carefully removed. The sample with RPE as a superficial layer was fixed in a holder system and covered with physiological saline solution. After irradiation, the sample was stained with the vitality marker CalceinAM (Molecular Probes). Due to the uncharged structure of CalceinAM, it can penetrate the cell membrane. Once inside the cell, the lipophilic blocking groups (AM) are cleaved by nonspecific esterases, and Calcein is unblocked. The intracellular released Calcein fluoresces when excited with blue light at 480nm .31 Living cells fluoresce brightly due to the accumulated Calcein compared to dead cells without esterases, which appear dark in the fluorescence microscope image (Fig. 2 ).

Fig. 2

CalceinAM stained porcine RPE cells. Live cells fluoresce while dead cells appear dark.

064022_1_013506jbo2.jpg

For the in vitro experiments, we used the clinical treatment parameter with 30 laser pulses at a repetition rate of 100Hz . Due to the lower damage threshold of in vitro porcine RPE,10 between 5to35μJ per pulse were applied to the samples.

2.2.

Patient Treatment

2.2.1.

Setup for patient treatment

We used the same laser/slitlamp system as described before. During treatment, a contact lens was placed on the patient’s cornea to eliminate the corneal refraction and to fix the eye. We modified a standard contact lens (Haag-Streit Laser-Lens 903L) with a piezzo-electric transducer (Fig. 3 ). The ring-shaped transducer (lead zirconium titanate, PZT5) with a resonance frequency of 1MHz was embedded into the contact lens and electromagnetically shielded by copper foil. The sensitivity of the OA contact lens was calibrated with a hydrophone (Ceram, number 118, 15mVbar ) to 5.1Vbar at 1MHz . For calibration, both detectors were placed in a water bath. They were adjusted to the same angle and the same distance to a black absorber sample. Laser pulses with 150ns in duration and 20to100-μJ pulse energy were applied to the sample, and the amplitude of the generated thermoelastic transient was measured simultaneously with an oscilloscope (TEK, TDS220).

Fig. 3

Setup for optoacoustic measurements during patient treatment with μs -laser pulses.

064022_1_013506jbo3.jpg

During treatment, the signals from the transducer were amplified (Panametrics, Model 5676, 40-dB amplification) and recorded by a data acquisition board (Datel, PCI 431-B, 10-MHz sample rate) in a personal computer (PC). The data acquisition process and the analysis were programmed with LabView (National Instruments 5.1). All treatments were recorded with a video system attached to the slitlamp to monitor the treatment and make the angiographic results comparable to the measured OA transients.

2.2.2.

Patient treatment parameters and proof of retinal pigment epithelium damage

The OA data were taken during patient treatment in a clinical pilot study for SRT at the eye clinic of the Medical University Lübeck. All 27 patients gave written informed consent to treatment and measurement. The clinical treatment parameters for SRT were 30 laser pulses at a repetition rate of 100Hz . Pulse energies ranged between 50 and 140μJ . Due to a magnification factor of 1.1 of the contact lens, the retinal illumination diameter was 176μm . Patients with diabetic maculopathy (DMP, 16 patients), central serous retinopathy (CSR, 5 patients), and drusen maculopathy (AMD, 6 patients) were treated. The number of laser lesions per treatment ranged from 10 up to 90 (mean 48) lesions per patient.

After treatments, fluorescein or indio-cyanine green (ICG) is intravenously injected. If RPE cells or their tight junctions are damaged, the outer blood-retina barrier is broken, and the dye can diffuse from the choriocapillaris to the retina. The damaged areas were visualized by fluorescein (FLA) or ICG angiography (Heidelberg Instruments, HRA).

2.3.

Acoustic Detection Simulation of the Optoacoustic Contact Lens

Due to the different treatment locations within the eye, the OA transients reach the piezo-electric transducer from different angles. This leads to strong distortions of the detected transients. The surgeon also tilts the contact lens during treatment to avoid dazzle of the illumination light reflection on the contact lens surface and to aim to different treatment locations. An adjustment to the acoustic axis of every treatment location during treatment is impractical; therefore, we studied the effect of acoustic transient distortion due to alteration of the acoustic path.

The effect on OA transient distortion due to different retinal treatment locations was examined by the use of an acoustic field simulation. The geometric assumptions of the simulations are shown in Fig. 4 . The software package ULTRASIM32 (programmed under MatLab at University of Oslo) was used. It simulates the acoustic field of various geometries under the Huygens principle, and was designed for optimization of medical ultrasonic transducers. A single sin2 -shaped 1-μs -long acoustic wave is emitted from 20 point sources within the 160-μm illumination diameter at different locations within the eye, as shown in Fig. 4. The pressure distribution within the detector plane is calculated for different time points. A numerical integration over the ring-shaped detector structure within the detector plane leads to the calculated transients.

Fig. 4

Geometry for acoustic simulation of the OA contact lens. The simulated retinal treatment locations range from 0to40deg .

064022_1_013506jbo4.jpg

2.4.

Optoacoustic Data Analysis

Although the measured OA transients showed a significant difference in pulse-to-pulse reproducibility for damaged and nondamaged RPE cells, the extraction of a reliable measure of this effect was not possible just by analyzing the frequency spectra or the acoustic power of the signals. The main problem was that, especially during patient treatment, the acoustic transfer function changed from treatment spot to treatment spot, and thus also the form and amplitude of the OA transients. Also, the subtraction of a thermoelastic reference transient is inappropriate. To overcome these problems, we developed an algorithm that takes these variabilities into account. The algorithm analyzes the OA transients in real time, and allows a determination if RPE cells are damaged or not.

In the first step, the mean transient P¯(t) of the 30 OA transients Pj(t) per irradiation area is calculated.

Eq. 1

P¯(t)=1nj=1nPj(t).
Second, the mean transient P¯(t) is subtracted from every single transient Pj(t) to get 30 deviation transients Dj(t) .

Eq. 2

Dj(t)=Pj(t)P¯(t).
After temporal integration in the time window t1 and t2 of the absolute value of the deviation transients Dj(t) , we extract the maximum of the n=30 integral values.

Eq. 3

OAvalue=max[t1t2Dj(t)dt]:j=1n.
This value, called the OA value, gives a measure of the maximum pressure difference between the OA transients within a pulse train.

Due to a lack of a thermoelastic reference transient during treatment, the mean transient P¯(t) from Eq. 1 is used. By subtraction from the measured transients Pj(t) , the thermoelastic parts are removed, as these are stable within one treatment spot. In the case of a subthreshold lesion, one extracts only the electrical and acoustic noise of the measurement. In a successful irradiation, microbubbles are generated. In this case, Eq. 2 extracts only the microbubble-induced acoustic transients, which are randomly formed. By choosing the maximum value [Eq. 3] of the deviation transients Dj(t) , the detection of a single microbubble formation is ensured.

3.

Results

3.1.

Detection Characteristic of the Optoacoustic Contact Lens

With respect to different retinal target points, the expansion of the OA transient was simulated as described in Sec. 2(C). The pressure distribution within the detector plane was calculated for different times after laser impact on the retinal location. Figure 5 shows a typical series of pressure distributions in the detector plane for a 10-deg tilt off the central spot. The decentral appearance of the pressure wave is clearly visible. For each temporal point, the numerical integration of the pressure values over the detector area leads to the pressure amplitude.

Fig. 5

Pressure distribution in the detector plane at different times for a 10-deg tilt of the target point in the eye. The decentral appearance of the pressure wave is clearly visible. For each temporal point, the numerical integration of the pressure values over the detector area leads to the pressure amplitude.

064022_1_013506jbo5.jpg

The simulated OA transients for the tilting range from 0to40deg are shown in Fig. 6 . During SRT, 0deg refers to a location slightly above the fovea, and 40deg will be outside of the retinal arcade vessels. The 0-deg transient shows the highest amplitude and shortest bipolar transient. Even in this case, the emitted sin2 burst is strongly distorted and stretched by a factor of 1.5, owing to the ring detector geometry. For higher tilt angles, the amplitude is lowered up to a factor of 2, and the transient is strongly distorted and stretched up to a factor of 5. This is due to the longer transmission time of the pressure wave over the detector area. The emitted mean frequency of 1MHz from the retinal target location is shifted to lower frequencies, down to 0.6MHz , while higher frequency components are strongly rejected.

Fig. 6

Simulated pressure transients according to the geometric tilt of the retinal target area. Depending on the tilting angle, the pressure transients are temporally stretched, showing different amplitudes and different temporal onsets.

064022_1_013506jbo6.jpg

With respect to these simulations, the variability of the expected amplitude, shape, and frequencies of the OA transient during SRT is expected to be high. Based on this simulation, the data analysis method described in Sec. 2(D) does not rely on a fixed OA transient shape.

3.2.

Optoacoustic Transients and Optoacoustic Retinal Pigment Epithelium Damage Detection In Vitro on Porcine Retinal Pigment Epithelium Samples

A typical set of different OA transients recorded during irradiation of RPE samples is shown in Fig. 7 . All graphs show a superposition of the 30 OA transients of a pulse train of different areas within one sample. At an exposure of 100mJcm2 , which is below the RPE damage threshold, 30 pure thermoelastic OA transients are measured [Fig. 7(a)]. As the distance of the acoustic detector to the sample was 2mm , the measurement was in the acoustic far field of the acoustic source ( 160-μm spot). For this reason, the OA transient is bipolar, followed by acoustic reflections within the detector. The 30 OA transients differ only by the noise level of the detection system. The measured pressure amplitude was 0.75mbar . Slightly above the RPE damage threshold [Fig. 7(b)], the bipolar parts of the 30 OA transients remained unchanged. Only at the end of the bipolar part, small pulse-to-pulse fluctuations with 100-μbar amplitude are detectable. Increasing the exposure 25% above the RPE damage threshold [Fig. 7(c)], we found that the OA pulse-to-pulse fluctuations were strongly increased and nearly as big as the thermoelastic part of the OA transients. Also, the temporal onset of the fluctuations is earlier compared to the lower exposures seen in Fig. 7(b). At 2.5 times the threshold exposure, the amplitudes of the fluctuations strongly overwhelmed the thermoelastic transient [Fig. 7(d)]. A second bipolar transient is generated after several microseconds. The delay between the first and the second bipolar peak is reduced for each of the following laser pulses.

Fig. 7

30 superimposed OA transients during irradiation of fresh porcine RPE. (a) Below RPE damage (100mJcm2) : only pure thermoelastic OA transients are observed, no pulse-to-pulse fluctuation. (b) Several damaged cells within the illuminated area (125mJcm2) : thermoelastic OA transient with small pulse-to-pulse fluctuations are seen. (c) 25% above the damage threshold (150mJcm2) : thermoelastic OA transients with strong pulse-to-pulse fluctuations and similar amplitudes are observed. (d) 250% over damage threshold (450mJcm2) : strong bubble formation transients are observed with the formation of a second bipolar bubble collapse transient; these are much higher pressure amplitudes compared with the pure thermoelastic transients.

064022_1_013506jbo7.jpg

All measured OA transients were analyzed as described in Sec. 2(D). In Fig. 8 , the OA value is plotted over the percentage of dead RPE cells within the illumination spot. The OA threshold value OAvitro=2.4×1010(bars) was determined by minimizing the number of false positive and false negative results. This value was now used to differentiate between damaging and nondamaging irradiations. It is therefore possible to detect a microbubble formation signal of several cells under the thermoelastic OA transients of 80 cells within the illumination area. Of the 142 illuminations of five RPE samples, only three were false positive and three false negative. The RPE damage threshold values were recalculated into their single dichotome values ( 0=dead , 1=live ) and analyzed with a Probit algorithm33, 34 (SPSS, Chicago IL). The damage threshold of porcine RPE cells was found to be ED50RPE=194mJcm2 with ED85RPE=251mJcm2 and ED15RPE=149mJcm2 as upper and lower widths of the adjusted logarithmic normal distribution, respectively. Converting the measured OA values to dichotome values ( 0:OAOAvitro , 1:OAOAvitro ) and utilizing the Probit analysis, a threshold for microbubble formation of ED50OA=92mJcm2 ( ED15OA=66mJcm2 ; ED85OA=128mJcm2 ) was determined.

Fig. 8

OA value versus percent RPE cell death of six samples of six different eyes in vitro. Over all, 142 areas were illuminated. The threshold for optoacoustically detected RPE cell damage was determined as OAvitro=2.4×10-10[bars] by minimizing the number of false-positive and false-negative results. In this case, only three locations were sorted as false positives and three as false negatives. The three false-negative results were still below the threshold for 50% cell death (ED50).

064022_1_013506jbo8.jpg

3.3.

Results During Selective Retina Treatment

During 27 patient (27 eyes) SRT treatments, the OA transients were recorded. Overall, 1370 lesions were placed.

3.3.1.

Optoacoustic transient during SRT patient treatment

All lesions placed in this study were ophthalmoscopically invisible. Due to the high sensitivity of 5.1Vbar of the OA contact lens, it was possible to measure the generated OA transients during SRT. Two typical OA transients are shown in Fig. 9 for an angiographic visible lesion. Three different areas of interest are marked. After an acoustic transmission time of 5μs , which corresponds to a geometric distance of 7.4mm , a thermoelastic transient is detected [Fig. 9(a)]. This part of the transient was only generated in patients with turbid eye structures. Pulse-to-pulse fluctuations were never observed within this time window (5to13μs) . This signal was never measured for the young patient group (average age 40years ) with CSR (compare Fig. 10 ), which had clear eye structures. The measured acoustic distance of 7.4mm of this OA transient generating structure relative to the OA contact lens corresponds to the distance of the anterior eye structures, i.e., the strong light absorbing iris where these signals are generated by the absorption of scattered light. After approximately 14μs , a thermoelastic bipolar transient was measured [Fig. 9(b)]. The acoustic delay time corresponded to a distance of 21mm (at 1532ms for vitreous body), which agrees with the distance of the transducer to the retina. This part in the time window between 14 and 22μs was found in all OA transients measured during SRT. If pulse-to-pulse fluctuations are detectable, their onset is within this time window. The bipolar peak was followed by acoustic reflections within the OA contact lens, and oscillation of only a slightly damped resonant transducer with its resonant frequency of 1MHz [Fig. 9(c)] was observed.

Fig. 9

Typical OA transients measured during SRT by use of the OA contact lens. Three different areas of interest are marked. (a) Thermoelastic transient induced by the anterior eye structure (iris) by scattered light from turbid eye media (was only detectable in patients with turbid media). (b) Thermoelastic transient of the RPE with pulse-to-pulse fluctuations. (c) Acoustic reflections within the OA contact lens and oscillation of the resonant transducer with its resonant frequency of 1MHz .

064022_1_013506jbo9.jpg

Fig. 10

Typical OA transients measured during patient treatment. (a) Subthreshold lesion at 50μJ , angiographic invisible lesion, no pulse-to-pulse fluctuations. (b) Suprathreshold lesion at 125μJ , angiographic visible lesion, small pulse-to-pulse fluctuations. (c) Suprathreshold lesion at 125μJ , angiographic visible lesion, strong pulse-to-pulse fluctuations with bubble collapse sometimes detectable (arrow).

064022_1_013506jbo10.jpg

Comparing OA transients of patients below and above angiographic visible lesions shows similar transients as in the in vitro experiments on porcine RPE samples. Three typical superpositions of the 30 OA transients at different energies and locations within one patient eye (CSR, age 28years ) are shown in Fig. 10. For exposures that result in no angiographic visible lesions (50μJ) , the 30 OA transients showed no pulse-to-pulse fluctuations [Fig. 10(a)], and a pure thermoelastic transient is measured. The measured pressure amplitudes are around 0.3mbar . At higher exposure [Fig. 10(b), 125μJ , angiographically visible], a strong thermoelastic bipolar transient is measured. However, small pulse-to-pulse fluctuations start within the bipolar peak. For the same exposure, but a different location within the eye, strong fluctuations and even a second bipolar peak is detectable for the first applied laser pulse [Fig. 10(c)]. This second bipolar peak was detected very rarely. In the over 1370 treatment spots that were applied during the 27 patient treatments, this second bipolar peak was detected only in three spots.

3.3.2.

Optoacoustic online dosimetry during patient treatment

Roider 6, 7, 8 introduced the fluorescein leakage observed in angiography as a reference standard for the detection of the ophthalmoscopic invisible selective RPE damage. With respect to our OA-based RPE damage detection, the OA values of every single spot were compared to the angiographic responses. A direct comparison of every single treatment spot was not possible in patients with DMP. The fundus of these patients shows a disease-related diffuse FLA leakage that makes it impossible to detect the additional laser-induced FLA spots. A direct correlation of treated laser spots succeeded only in the small group of four CSR patients. Nearly every single treatment lesion could directly be compared to the FLA result.

As the OA transients during patient treatment showed similar characteristics in comparison to the in vitro data, the same analysis algorithm was used. This analysis was limited to a temporal window from t1=12μs to t2=30μs , where most of the fluctuations appeared. To reduce the electromagnetic noise, the OA transients were filtered digitally with a fourth-order Butterworth filter (LabView) between 5kHz and 1MHz .

In Fig. 11 , all analyzed OA values of these four treatments are plotted with respect to the pulse energy. The open symbols mark all angiographic visible lesions, filled gray symbols are the FLA negative, and the filled black symbols are the data points that were not analyzable due to multiple lesions being placed on the same spot. The Probit algorithm33, 34 was used to determine the OA threshold value in vivo as OAvivo=1.96×1010[bars] . In these four treatments, two out of 94 lesions were detected as false positives and two as false negatives. The OA threshold for an angiographic lesion during patient treatment of OAvivo=1.96×1010[bars] was found to be close to the threshold for porcine RPE damage, OAvitro=2.4×1010[bars] .

Fig. 11

OA value versus applied pulse energy for four patient treatments. The open symbols mark the FLA positive lesions, gray symbols mark the FLA negative, and the black symbols were not analyzable data points.

064022_1_013506jbo11.jpg

As seen in Fig. 11, the OA values vary a lot within one eye at the same applied energy. In patients with DMP, lower OA values were generally achieved in edematous retinal locations.

3.3.3.

Optoacoustic value in comparison with indigo-cyanine green angiography intensity

In the in vitro results of Fig. 8, it can be seen that as the OA values increased, more RPE cells were damaged within the treatment spot. During patient treatment, the amount of damaged RPE cells within a single spot was not accessible. During fluorescein angiography, the diffusion of fluorescein through the damaged RPE areas was too fast to get a measurement of the leakage size. The clearest FLA images are always seen in the early phase, several seconds after dye injection. In contrast to this, ICG diffuses very slowly through the damaged RPE areas. The best contrast for the visualization was in the very late phase, up to 20min after injection. In ICG angiograms, the single treatment spots showed different fluorescence intensity [Fig. 12(a) ]. Due to the nonlinearity of the whole ICG angiography process, the mean gray value of the spot gave a rough hint of the amount of damaged RPE cells within that spot.

Fig. 12

(a) ICG angiography image with lesion locations (white circles) and the associated lesion number. (b) Analyzed OA values over mean gray value of the corresponding ICG spots. The brighter the ICG spot, the higher the measured OA values.

064022_1_013506jbo12.jpg

The ICG angiogram image in Fig. 12(a) shows the location of each single lesion as a white circle. Within these circles, the mean gray value was analyzed for each associated lesion number. These mean gray values of the ICG angiogram are plotted with respect to the OA value for every corresponding spot number [Fig. 12(b)]. It can be seen that the brighter the ICG spot, the higher the corresponding OA value. From this data, it appeared that during patient treatment, the OA dosimetry system detects the amount of damage applied to the RPE cells.

3.3.4.

Optoacoustic classification of 27 SRTs

Overall, the OA dosimetry system was used during 27 SRT patient treatments. In the first stage of this study, test lesions at various laser pulse energies were applied to the fundus around the retinal vessel arcades. Based on the immediate fluorescein angiography results, the treatment in the central region was performed with laser pulse energies that showed successful FLA leakage in the test lesions. As the measured OA values during central treatment were far higher than the test lesions, the treatment energy was lowered according to the OA values obtained. Final FLAs after complete treatment indicated RPE damage in the central region at lower pulse energies, adjusted due to the OA value.

In the later phase of the study, the first FLA for the test lesion was canceled, and the treatment energy in the central area was only adjusted based on the OA values. In all cases, the FLA after laser treatment indicated successful damage of RPE, as noted by fluorescein leaking at the treatment spots.

To obtain an overview of the data for all 27 treatments, the results of the 1370 spots are summarized in Fig. 13 . The measured OA values are sorted into three different categories. First, the range of OA values below the threshold of OAvivo=1.96×1010[bars] , with no pulse-to-pulse fluctuations and no RPE damage. Second, the therapeutic range of OA values are between 1.96×1010[bars] and 8×1010[bars] , with angiographicly visible RPE damage and small pulse-to-pulse fluctuations. Finally, the range of OA values larger than 8×1010[bars] is grouped, with strong pulse-to-pulse fluctuations and occasionally detectable bubble collapse.

Fig. 13

Probability of three OA value ranges over pulse energies for 27 patient treatments ( n=1370 lesions).

064022_1_013506jbo13.jpg

During the treatments, the laser pulse energy was adjusted so that the OA values were within the range of successful therapeutic treatment. This was difficult, even within one eye. On this basis, 23% of the lesions at 100-μJ pulse energies were overtreated, while 20% still did not show an adequate OA response. The variability of OA values over the therapeutic energy range from 80to150μJ clearly demonstrates the necessity of a real-time dosimetry system during SRT.

4.

Discussion

In this work, we introduced and proved the efficiency of an OA based real-time dosimetry system for SRT during in vitro and in vivo treatment.

4.1.

Detection of Optoacoustic Transients In Vivo and During Patient Treatment

Due to the high sensitivity of the 1.05-Vbar ultrasonic transducer used for the in vitro experiments and the broadband preamplifier ( 54-dB amplification), we were able to reduce the noise level of the pressure amplitude down to 10μbar at 1MHz (Fig. 7). This high sensitivity enables the detection of thermoelastic transients during μs irradiation, which far exceeds the acoustic confinement time of the applied laser pulse, so that the pressure generation is less effective. As microbubble formation around the strong absorbing melanosomes inside the RPE cells was shown to be the primary RPE damage mechanism10, 17 for μs laser pulses, the onset of the detected pulse-to-pulse fluctuations of the OA transients [as seen in Fig. 7(b)] was consistent with the onset of damaged RPE cells within the exposed spot. These fluctuations can be linked to the acoustic formation and collapse transients during microbubble formation.29 As the RPE cells contain several hundred melanosomes, and 80 RPE cells are illuminated within one spot, irregular microbubble formation is most probable. Also, early onset of the fluctuations by increasing the exposure [Figs. 7(b), 7(c), 7(d)] can be explained by an earlier microbubble formation due to effective heating.35 The second bipolar transients in Fig. 7(d) are most likely the collapse transients of big “macrobubbles,” which were also detected by Kelly during threefold threshold exposures using nanosecond laser pulses.19 The reduction of the delay time between the two following bipolar peaks can be induced by material ejection of the previous laser pulse. The OA detection system works well and reliably on porcine RPE samples. Even the damage of several cells within a spot of 80 illuminated cells can be detected. Therefore, the threshold ED50OA for microbubble formation is below the RPE damage threshold.

Due to the high sensitivity of 5.1Vbar at 1MHz of the OA contact lens, it was possible to detect the OA transients during patient treatment. The noise level could be reduced down to 2μbar (Fig. 9). The OA transients measured during patient treatment coincided with the OA transients from experiments on porcine RPE samples. If RPE damage is detectable via an angiographic visible lesion in patients, then pulse-to-pulse fluctuations appear at the end of the thermoelastic bipolar peak [Fig. 10(b)]. Similar pulse-to-pulse fluctuations were detected if porcine RPE cells were damaged, as indicated by vitality staining with calceinAM. Also, in patient treatment, the fluctuations are linked to the microbubble formation around the melanosomes inside the RPE cells. This is further supported by the detection of a second bipolar peak [Fig. 10(c)], which can be identified as the bubble collapse transient.30 In this case, the bubble lifetime was 6μs .

The results of the simulation of the OA contact lens characteristics (Figs. 5 and 6) show that the acoustic transfer function strongly changes for different locations within the eye. During patient treatment, it was not possible to align the contact lens well at each treatment location, thus measurements of absolute pressure amplitudes and waveforms were not possible. This was taken into account by the analysis method used (Sec. 2(D)). The method used the mean transient P¯(t) of all OA transients per pulse train as a reference, as the eye will not significantly move during the 300-ms exposure time. At every single spot, a new reference transient was defined and changes in the acoustic transfer function, for example due to movements, were eliminated. The developed analysis method is fast ( 30.2ms on a Pentium 120MHz ), which makes it applicable for a real-time dosimetry system.

4.2.

Optoacoustic Dosimetry System Sensitivity

Until now, selective RPE damage can only be detected by fluorescein or ICG angiography.6, 7, 8, 36 For this procedure, the dye is intravenously injected into the patient and visualized at the fundus with a fundus camera or laser scanning angiograph. Besides its invasive character, the main drawback of this monitoring technique is that it cannot be used as a real-time dosimetry method, which leads to practical limitations of SRT. For an ideal treatment, several steps are needed to insure a successful irradiation. First, test lesions at different energy levels are applied to the fundus around the retinal vessel arcades, and the first angiogram depicts the threshold energy. Second, the treatment is then performed with the determined threshold energy; however, a second angiogram is needed to confirm the induced lesions. In some cases, a second treatment is needed if the initially delivered treatment energy was too low. This invasive and time-consuming procedure is a practical limitation for SRT to become a standard treatment modality that can be performed in private practice. A real-time dosimetry method is highly desirable to bring this novel technique to broad application.

Our in vitro results show that the OA detection method is sensitive enough to detect only several damaged RPE cells within an exposed spot of 80 cells. An acoustically detected damage threshold can be detected at OAvitro=2.4×1010(bars) (Fig. 8). Due to the high acoustic detection sensitivity, the RPE damage threshold value of ED50RPE=194mJcm2 (50% RPE damage) is higher than the threshold for microbubble formation ED50OA=92mJcm2 (50% occurrence of microbubble formation within the spot). This also indicates that microbubble formation can be clearly detected if less than 50% of cells are damaged. In a more recent study,17 we could also show that microbubble formation can even be detected if only one RPE cell is damaged. During patient treatment, an acoustic detected damage threshold can be defined as OAvivo=1.96×1010(bars) (Fig. 11). This value is close to the threshold value for in vitro experiments (OAvitro) , which is surprising due to the different geometries of the transducers and acoustic transfer functions, which were not taken into account.

The results of the in vitro experiments (Fig. 8) also show that the higher the extracted OA value was, the more cells were damaged within one spot. The amount of damaged RPE cells cannot be directly analyzed in a patient. Therefore, as a very rough estimate, the gray level of an ICG angiography was correlated to the amplitude to the measured AO value. In Fig. 12(b) it can be seen that the brighter the ICG spot was, the higher the corresponding OA value. From this data, it seems that during patient treatment, the OA dosimetry system detects the amount of damaged RPE cells. However, a determination of the macroscopic damage was not possible. A better correlation of the detected OA value and the extent of RPE damage can only be done in direct comparison of single spots under histological examination, which is not possible in humans. Ultra-high-resolution optical coherence tomography (OCT) images,37 which are applicable in humans, might allow further insight.

Due to the high variability in pulse energy that is needed to reach the acoustic threshold OAvivo during patient treatments (Fig. 13), it appeared that an active control of the laser pulse energy would be reasonable. Especially in patients with diabethic maculopathy and hard drusen, the threshold energy was strongly varying, depending on the treatment location.

5.

Conclusion

In this work, we introduce and prove the concept of an OA-based real-time dosimetry system for SRT. In porcine RPE samples in vitro, as well as in patient treatments, the OA dosimetry system detected the laser-induced RPE damage. More than 27 patient treatments are controlled by this system. The OA dosimetry system is embedded in a clinical SRT prototype laser and will be further evaluated during an ongoing clinical multicenter SRT study.

References

1. 

Macular Photocoagulation Study Group, “Laser photocoagulation of subfoveal recurrent neovascular lesions in age-related macular degeneration. Results of a randomized clinical trial.,” Arch. Ophthalmol. (Chicago), 109 (9), 1232 –1241 (1991). 0003-9950 Google Scholar

2. 

Macular Photocoagulation Study Group, “Argon laser photocoagulation for senile macular degeneration. Results of a randomized clinical trial,” Arch. Ophthalmol. (Chicago), 100 (6), 912 –918 (1982). 0003-9950 Google Scholar

3. 

The Diabetic Retinopathy Study Research Group, “Photocoagulation treatment of proliferative diabetic retinopathy: relationship of adverse treatment effects to retinopathy severity. Diabetic retinopathy study report no. 5,” Dev. Ophthalmol., 2 248 –261 (1981). 0250-3751 Google Scholar

4. 

Treatment of Age-related Macular Degeneration with Photodynamic Therapy (TAP) Study Group, “Photodynamic therapy of subfoveal choroidal neovascularization in age-related macular degeneration with verteporfin: one-year results of 2 randomized clinical trials—TAP report.,” Arch. Ophthalmol. (Chicago), 117 (10), 1329 –1345 (1999). 0003-9950 Google Scholar

5. 

M. A. Mainster and E. Reichel, “Transpupillary thermotherapy for age-related macular degeneration: long-pulse photocoagulation, apoptosis, and heat shock proteins,” Ophthalmic Surg. Lasers, 31 (5), 359 –373 (2000). 1082-3069 Google Scholar

6. 

J. Roider, R. Brinkmann, C. Wirbelauer, H. Laqua, and R. Birngruber, “Subthreshold (retinal pigment epithelium) photocoagulation in macular diseases: a pilot study,” Br. J. Ophthamol., 84 (1), 40 –47 (2000). https://doi.org/10.1136/bjo.84.1.40 0007-1161 Google Scholar

7. 

J. Roider, F. Hillenkamp, T. Flotte, and R. Birngruber, “Microphotocoagulation: Selective effects of repetitive short laser pulses,” Proc. Natl. Acad. Sci. U.S.A., 90 8643 –8647 (1993). 0027-8424 Google Scholar

8. 

J. Roider, R. Brinkmann, C. Wirbelauer, H. Laqua, and R. Birngruber, “Retinal sparing by selective retinal pigment epithelial photocoagulation,” Arch. Ophthalmol. (Chicago), 117 (8), 1028 –1034 (1999). 0003-9950 Google Scholar

9. 

B. R. Hammond and M. Caruso-Avery, “Macular pigment optical density in a southwestern sample,” Invest. Ophthalmol. Visual Sci., 41 (6), 1492 –1497 (2000). 0146-0404 Google Scholar

10. 

R. Brinkmann, G. Huettmann, J. Rogener, J. Roider, R. Birngruber, and C. P. Lin, “Origin of retinal pigment epithelium cell damage by pulsed laser irradiance in the nanosecond to microsecond time regimen,” Lasers Surg. Med., 27 (5), 451 –464 (2000). https://doi.org/10.1002/1096-9101(2000)27:5<451::AID-LSM1006>3.0.CO;2-1 0196-8092 Google Scholar

11. 

R. Klein, B. E. Klein, and S. E. Moss, “Epidemiology of proliferative diabetic retinopathy,” Diabetes Care, 15 (12), 1875 –1891 (1992). 0149-5992 Google Scholar

12. 

F. L. Ferris, “How effective are treatments for diabetic retinopathy,” JAMA, J. Am. Med. Assoc., 269 (10), 1290 –1291 (1993). https://doi.org/10.1001/jama.269.10.1290 0098-7484 Google Scholar

13. 

J. D. Gass, Stereoscopic Atlas of Macular Disease: Diagnosis and Treatment, (1997) Google Scholar

14. 

H. M. Leibowitz, D. E. Krueger, L. R. Maunder, R. C. Milton, M. M. Kini, H. A. Kahn, R. J. Nickerson, J. Pool, T. L. Colton, J. P. Ganley, J. I. Loewenstein, and T. R. Dawber, “The Framingham eye study monograph: An ophthalmological and epidemiological study of cataract, glaucoma, diabetic retinopathy, macular degeneration, and visual acuity in a general population of 2631 adults, 1973–1975,” Surv. Ophthalmol., 24 335 –610 (1980). https://doi.org/10.1016/0039-6257(80)90015-6 0039-6257 Google Scholar

15. 

R. Klein, B. E. Klein, S. C. Jensen, and S. M. Meuer, “The five-year incidence and progression of age-related maculopathy: the beaver dam eye study,” Ophthalmology, 104 (1), 7 –21 (1997). 0161-6420 Google Scholar

16. 

B. Rahmani, J. M. Tielsch, J. Katz, J. Gottsch, H. Quigley, J. Javitt, and A. Sommer, “The cause-specific prevalence of visual impairment in an urban population. The Baltimore eye survey,” Ophthalmology, 103 (11), 1721 –1726 (1996). 0161-6420 Google Scholar

17. 

G. Schuele, M. Rumohr, G. Huettmann, and R. Brinkmann, “RPE damage thresholds and mechanisms for laser exposure in the μstoms time regimen,” Invest. Ophthalmol. Visual Sci., 46 714 –719 (2005). https://doi.org/10.1167/iovs.04-0136 0146-0404 Google Scholar

18. 

C. P. Lin and M. W. Kelly, “Cavitation and emission around laser-heated microparticles,” Appl. Phys. Lett., 72 (22), 1 –3 (1998). https://doi.org/10.1063/1.121437 0003-6951 Google Scholar

19. 

M. W. Kelly, “Intracellular cavitation as a mechanism of short-pulse laser injury to the retinal pigment epithelium,” Electrical Engineering, (1997) Google Scholar

20. 

R. Birngruber, “Thermal modelling in biological tissue,” Lasers in Biology and Medicine, 77 –97 Plenum, New York (1980). Google Scholar

21. 

R. Birngruber, F. Hillenkamp, and V. P. Gabel, “Theoretical investigations of laser thermal retinal injury,” Health Phys., 48 (6), 781 –796 (1985). 0017-9078 Google Scholar

22. 

M. W. Sigrist and F. K. Kneubühl, “Laser-generated stress waves in liquids,” J. Acoust. Soc. Am., 64 (6), 1652 –1663 (1978). https://doi.org/10.1121/1.382132 0001-4966 Google Scholar

23. 

M. W. Sigrist, “Laser generation of acoustic waves in liquids and gases,” J. Appl. Phys., 60 (7), R83 –R121 (1986). https://doi.org/10.1063/1.337089 0021-8979 Google Scholar

24. 

G. Schuele, G. Huettmann, C. Framme, J. Roider, and R. Brinkmann, “Noninvasive optoacoustic temperature determination at the fundus of the eye during laser irradiation,” J. Biomed. Opt., 9 (1), 173 –179 (2004). https://doi.org/10.1117/1.1627338 1083-3668 Google Scholar

25. 

U. Oberheide, I. Bruder, H. Welling, W. Ertmer, and H. Lubatschowski, “Optoacoustic imaging for optimization of laser cyclophotocoagulation,” J. Biomed. Opt., 8 (2), 281 –287 (2003). https://doi.org/10.1117/1.1559998 1083-3668 Google Scholar

26. 

R. O. Esenaliev, I. V. Larina, K. V. Larin, and M. Motamedi, “Real-time optoacoustic monitoring during thermotherapy,” Proc. SPIE, 3916 302 –310 (2000). https://doi.org/10.1117/12.386334 0277-786X Google Scholar

27. 

R. O. Esenaliev, A. A. Oraevsky, K. V. Larin, I. V. Larina, and M. Motamedi, “Real-time optoacoustic monitoring of temperature in tissues,” Proc. SPIE, 3601 268 –275 (1999). https://doi.org/10.1117/12.350008 0277-786X Google Scholar

28. 

K. V. Larin, I. V. Larina, M. Motamedi, and R. O. Esenaliev, “Monitoring of temperature distribution in tissues with optoacoustic technique in real time,” Proc. SPIE, 3916 311 –321 (2000). https://doi.org/10.1117/12.386336 0277-786X Google Scholar

29. 

G. Schuele, G. Huettmann, J. Roider, C. Wirbelauer, R. Birngruber, and R. Brinkmann, “Optoacoustic measurements during μs-s irradiation of the retinal pigment epithelium,” Proc. SPIE, 3914 230 –236 (2000). https://doi.org/10.1117/12.388090 0277-786X Google Scholar

30. 

A. Vogel and S. Busch, “Schock wave emission and cavitation bubble generation by picosecond and nanosecond optical breakdown in water,” J. Acoust. Soc. Am., 100 (1), 148 –165 (1996). https://doi.org/10.1121/1.415878 0001-4966 Google Scholar

31. 

Molecular Probes, “LIVE/DEAD viability/cytotoxicity,” (2001) Google Scholar

32. 

S. Holm, F. Teigen, L. Odegaard, V. Berre, and J. O. Erstad, ULTRASIM User’s Manual, Univ. of Oslo(1998). Google Scholar

33. 

, Technical Support, Probit ( (2001) Google Scholar

34. 

D. J. Finney, “Probit analysis: statistical treatment of the sigmoid response curve,” (1964) Google Scholar

35. 

J. Roegener, R. Brinkmann, and C. P. Lin, “Pump-probe detection of laser-induced microbubble formation in retinal pigment epithelium cells,” J. Biomed. Opt., 9 (2), 367 –371 (2004). https://doi.org/10.1117/1.1646413 1083-3668 Google Scholar

36. 

J. Roider, C. Lindemann, E. S. El-Hifnawi, H. Laqua, and R. Birngruber, “Therapeutic range of repetitive nanosecound laser exposures in selective RPE photocoagulation,” Graefe's Arch. Clin. Exp. Ophthalmol., 236 213 –219 (1998). 0721-832X Google Scholar

37. 

J. G. Fujimoto, “Optical coherence tomography for ultrahigh resolution in vivo imaging,” Nat. Biotechnol., 21 (11), 1361 –1367 (2003). https://doi.org/10.1038/nbt892 1087-0156 Google Scholar
©(2005) Society of Photo-Optical Instrumentation Engineers (SPIE)
Georg Schuele, Hanno Elsner, Carsten Framme, Johann Roider M.D., Reginald Birngruber, and Ralf Brinkmann "Optoacoustic real-time dosimetry for selective retina treatment," Journal of Biomedical Optics 10(6), 064022 (1 November 2005). https://doi.org/10.1117/1.2136327
Published: 1 November 2005
Lens.org Logo
CITATIONS
Cited by 66 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Angiography

Acoustics

In vitro testing

Pulsed laser operation

Retina

Eye

Transducers

Back to Top